Management of Lost Atherectomy Devices in the Coronary Arteries
Gregor Leibundgut1, Alexandru Achim2, Daniel Weilenmann3
1Department of Cardiology, University Hospital Basel, Basel, Switzerland.
Insights
Lost atherectomy devices are a rare but serious complication during coronary procedures. This review updates management strategies, including a novel conservative approach for lost rotational or orbital atherectomy devices.
Area of Science:
- Interventional Cardiology
- Cardiovascular Devices
- Medical Complications
Background:
- Rotational atherectomy and orbital atherectomy are key treatments for calcific coronary disease.
- Device entrapment and subsequent loss are serious complications during these procedures.
- Increased adoption of atherectomy necessitates understanding and managing these rare events.
Purpose of the Study:
- To review contemporary management strategies for lost atherectomy devices in coronary arteries.
- To present a unique cohort of patients with retained atherectomy devices.
- To propose an updated algorithmic approach to managing lost atherectomy devices.
Main Methods:
- Review of literature and case series on lost atherectomy devices.
- Analysis of management techniques for device entrapment and loss.
- Development of a management algorithm incorporating conservative strategies.
Main Results:
- Current recommendations for entrapment typically involve device removal via percutaneous or surgical methods.
- Conservative management of retained devices has not been traditionally recommended.
- The study introduces a contemporary algorithmic approach, including a conservative arm.
Conclusions:
- Management of lost atherectomy devices lacks a clear consensus due to limited reported cases.
- An updated algorithmic approach, considering a conservative strategy, is proposed.
- This review aims to guide interventional cardiologists in managing this challenging complication.
Abstract:
Rotational and orbital atherectomy are important tools to treat calcific coronary disease. Entrapment of an atherectomy device, that is, rotational atherectomy burr or orbital atherectomy crown, is a serious complication during atherectomy. Loss of an atherectomy device is a more challenging complication that usually follows device entrapment. Although uncommon, given the increased adoption of atherectomy, it is important for interventional cardiologist to understand the mechanisms and principles for prevention and management of lost atherectomy devices. Given the paucity of reported cases, there is no clear consensus or defined approach as to how this complication is best managed. The current recommendations for burr/crown entrapment are generally to remove everything from the coronary vasculature and range from different percutaneous strategies and ultimately cardiac surgery. Conservative treatment with leaving the burr within the coronary vasculature has not been recommended. This article serves to review and provide a contemporary update on the management of lost atherectomy devices in the coronary arteries. We review the management and techniques centered around cases of lost atherectomy devices including a unique cohort of patients with lost burrs/crowns that were left in the coronary arteries. Finally, we provide an algorithmic approach to the contemporary management of lost atherectomy devices, incorporating a conservative strategy arm, and discuss situations where this may be considered.