Polarization diagnosis and transmission distortion calibration of the Dalian coherent light source
Mingchang Wang1,2, Bingbing Zhang3, Siyuan Tan1
1State Key Laboratory of Chemical Reaction Dynamics and Dalian Coherent Light Source, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, 457 Zhongshan Road, Dalian 116023, China.
Abstract:
Polarized light serves as a fundamental tool in various scientific research, particularly in the study of chiral molecules and magnetic materials. For Free-electron Laser (FEL) sources, the Elliptical Polarized Undulator (EPU) has been extensively used as a tool to generate tunable polarized FEL. The accurate adjustment of the polarization state critically depends on the precise diagnosis of the polarization state. This work presents a Vacuum Ultraviolet (VUV) polarimeter that employs a reflection polarizer to determine the polarization state of the Dalian Coherent Light Source (DCLS), a VUV-FEL facility in Dalian, China. This instrument has been precisely installed on the DCLS-FEL2 beamline to measure the EPU-generated FEL pulses. The linear polarization angle was measured with an accuracy of ±1°, and a polarization degree of 95% was achieved at 10.2 eV photon energy. Furthermore, beamline transmission matrix calibration has been conducted to ensure precision of polarization state at downstream end-stations.