Clinical and Radiological Results of the Integrated Acetabular MUTARS® RS Cup System in the Revision of Geriatric
Florian Radetzki1, Pawel Strzelczyk2, Horia Simion3
1Department of Orthopedic and Trauma Surgery, Dessau Medical Center, Brandenburg Medical School Theodor Fontane, Dessau-Roßlau, Germany.
Introduction:
Acetabular revision surgery is challenging due to the occurrence of bony defects that complicate the fixation of implants after removal of loose components. In addition to multimorbidity and increased risk of falling, poor bone quality in geriatric patients makes it even more difficult to secure implant anchoring. The aim of this study was to evaluate the clinical and radiographic results after revision arthroplasty with the MUTARS® RS cup system (implantcast GmbH, Buxtehude, Germany) in geriatric patients (age >70 years).
Materials And Methods:
From 2019 to 2023, 24 geriatric patients, six males and 18 females, with a mean age of 79.2 years (range, 70-94 years) underwent a revision surgery and were provided with the MUTARS® RS cup system at the geriatric center of the Dessau Municipal Hospital. After a mean follow up of 2.5 years (range, 1-4 years), the clinical evaluation and patient-reported outcomes were measured using the Harris Hip Score (HHS), Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC), Oxford Hip Score (OHS), and EQ-5D-5L. Radiographs were taken to evaluate radiolucent lines and component migration.
Results:
A total of 19 patients (79.2%) remained without any revision until the latest follow up. The average HHS was 71.1 (range, 56-93), the average OHS was 23.7 (range, 10-40), and the average WOMAC was 20.4 (range, 1-70). According to EQ-5D-5L, there were 10 patients (52.6%) with slight problems, seven patients (36.8%) with moderate problems, and two patients (10.5%) with severe problems. The x-ray showed no migration of the components; 15 cases (78.9%) had radiolucent lines around the caudal flap in the os ischii.
Conclusion:
The RS Cup is a suitable implant for revision surgery with large bone defects or acetabular fractures. "Clinically satisfactory" to "good" results are shown, with mild and moderate problems for geriatric patients which are already predominantly constrained by pre-existing multimorbidity.
More Related Videos
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
08:40Author Spotlight: A Novel 3D-Printed Titanium Implant for Minimally Invasive Treatment of Hip Dysplasia in Young Dogs
Published on: April 19, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Imaging Studies IV: Magnetic Resonance Imaging