Point-of-Care Ultrasound Versus X-ray for Pediatric Upper Extremity Injuries: A Safe and Radiation-Free Alternative?
Ioannis Delniotis1,2,3, Vangelis Bontinis2, Alexandros Delniotis1
1Department of Orthopedic, Hippokration General Hospital, Thessaloniki, Greece - Specialised Pediatric Orthopedics Practice.
Objectives:
This study aims to compare the diagnostic accuracy of point-of-care ultrasound (POCUS) and x-ray in detecting upper extremity fractures in children, focusing on sensitivity, specificity, and overall diagnostic performance. Particular attention was given to distal forearm fractures, the most common fracture site in children.
Methods:
This prospective diagnostic accuracy study was conducted in 2024 at a pediatric orthopedic emergency department. A cohort of 106 children aged 0 to 16 years, presenting with suspected upper extremity fractures, was enrolled. Each participant underwent both POCUS and x-ray imaging within a 24-hour period. POCUS was performed by a clinician, while x-ray images were interpreted by radiologists. The study evaluated sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV), and likelihood ratios for both general upper extremity fractures and specifically for distal forearm fractures.
Results:
Among the 106 participants, POCUS demonstrated an overall sensitivity of 88.3% (95% CI: 77.4-95.2) and specificity of 87.0% (95% CI: 73.7-95.1). The PPV was 89.8% (95% CI: 79.2-96.2), and the NPV was 85.1% (95% CI: 71.7-93.8). In the subgroup analysis of distal forearm fractures, POCUS exhibited enhanced diagnostic performance, with a sensitivity of 96.9% (95% CI: 83.8-99.9) and specificity of 93.3% (95% CI: 68.1-99.8). The PPV and NPV for distal forearm fractures were 96.9% (95% CI: 83.8-99.9) and 93.3% (95% CI: 68.1-99.8), respectively.
Conclusions:
POCUS proves to be a reliable, radiation-free diagnostic tool for detecting upper extremity fractures in children, particularly distal forearm fractures. However, due to the joint's complex anatomy, caution is advised when using POCUS for elbow fractures, which may increase the risk of misinterpretation.
More Related Videos
06:15Protocol and Guidelines for Point-of-Care Lung Ultrasound in Diagnosing Neonatal Pulmonary Diseases Based on International Expert Consensus
Published on: March 6, 2019
07:12Author Spotlight: Using Point-of-Care Ultrasound for Comprehensive Evaluation of the Abdominal Aorta
Published on: September 8, 2023
Related Concept Videos
X-ray Imaging
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...