Effect of data-driven motion correction for respiratory movement on lesion detectability in PET-CT: a phantom study
Marloes A de Winter1, Robin Gevers2, Jules Lavalaye3
1Department of Medical Physics, St Antonius Hospital, Nieuwegein, The Netherlands. ma.de.winter1@antoniusziekenhuis.nl.
Purpose:
While data-driven motion correction (DDMC) techniques have proven to enhance the visibility of lesions affected by motion, their impact on overall detectability remains unclear. This study investigates whether DDMC improves lesion detectability in PET-CT using FDG-18F.
Method:
A moving platform simulated respiratory motion in a NEMA-IEC body phantom with varying amplitudes (0, 7, 10, 20, 30 mm) and target-to-background ratios (2, 5, 10.5). Scans were reconstructed with and without DDMC, and the spherical targets' maximal and mean recovery coefficient (RC) and contrast-to-noise ratio (CNR) were measured.
Results:
DDMC results in higher RC values in the target spheres. CNR values increase for small, high-motion affected targets but decrease for larger spheres with smaller amplitudes. A sub-analysis shows that DDMC increases the contrast of the sphere along with a 36% increase in background noise.
Conclusion:
While DDMC significantly enhances contrast (RC), its impact on detectability (CNR) is less profound due to increased background noise. CNR improves for small targets with high motion amplitude, potentially enhancing the detectability of low-uptake lesions. Given that the increased background noise may reduce detectability for targets unaffected by motion, we suggest that DDMC reconstructions are used best in addition to non-DDMC reconstructions.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System V: CT