Related Experiment Video
Updated: Sep 16, 2025

Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
Impact of heart rate on coronary artery stenosis grading accuracy using deep learning-based fast kV-switching CT: A
Ryoji Mikayama1, Tsukasa Kojima2, Takashi Shirasaka3
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan; Department of Health Sciences, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan.
Introduction:
Deep learning-based fast kV-switching CT (DL-FKSCT) generates complete sinograms for fast kV-switching dual-energy CT (DECT) scans by using a trained neural network to restore missing views. Such restoration significantly enhances the image quality of coronary CT angiography (CCTA), and the allowable heart rate (HR) may vary between DECT and single-energy CT (SECT). This study aimed to examine HR's effect onCCTA using DL-FKSCT.
Methods:
We scanned stenotic coronary artery phantoms attached to a pulsating cardiac phantom with DECT and SECT modes on a DL-FKSCT scanner. The phantom unit was operated with simulated HRs ranging from 0 (static) to 50-70 beats per minute (bpm). The sharpness and stenosis ratio of the coronary model were quantitatively compared between DECT and SECT, stratified by simulated HR settings using the paired t-test (significance was set at p < 0.01 with a Bonferroni adjustment for multiple comparisons).
Results:
Regarding image sharpness, DECT showed significant superiority over SECT. In terms of the stenosis ratio compared to a static image reference, 70 keV virtual monochromatic image in DECT exhibited errors exceeding 10 % at HRs surpassing 65 bpm (p < 0.01), whereas 120 kVp SECT registered errors below 10 % across all HR settings, with no significant differences observed.
Conclusions:
In DL-FKSCT, DECT exhibited a lower upper limit of HR than SECT. Therefore, HR control is important for DECT scans in DL-FKSCT.
More Related Videos
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT