Related Experiment Video
Updated: Sep 15, 2025

Scanning Electron Microscopic Evaluation of Surface Defects of Remover Retreatment File After Single and Multiple Uses
Published on: October 11, 2024
Safety, cost and environmental impact of reprocessing high risk single-use medical devices: a systematic review and
Niamh McGrath1, Catherine Waldron1, Ailish Farragher1
1Evidence Centre, Health Information and Evidence Directorate, Health Research Board, Dublin, Ireland.
Aim:
To estimate the safety, financial and environmental effects of reprocessing high risk SUMDs.
Methods:
Systematic review (PROSPERO ID: CRD42022365642) of primary trial and observational studies of human participants receiving reprocessed high risk SUMDs compared with first use of identical SUMDs. Reprocessing was defined as cleaning, disinfection, and sterilisation or related procedures, and function and safety testing. Items were sourced via database, grey literature and supplemental searching of English and German language sources. Included studies were quality appraised and primary outcomes (direct patient safety; indirect financial costs; environmental impacts) GRADE (Grade of Recommendation, Assessment, Development and Evaluation) assessed. Narrative synthesis and where feasible, meta-analysis were undertaken.
Results:
Ten studies (N=2,657 participants) examined two implantable (pacemaker, defibrillator) and three catheterisation (electrophysiology polyurethane, ablation and balloon) devices. Safety outcomes were available for both device types and cost outcomes were available for catheterisation devices. Except for one older study, there were no statistically significant differences in the odds of examined safety outcomes between new and once reprocessed SUMDs. Meta-analysis of catheterisation devices resulted in similar results (Infections: OR=0.67, 95% CI: 0.37-1.20, p=0.18; Battery depletion: OR=0.2.29, 95% CI: 0.83-6.31, p=0.11). One study of balloon catheterisation devices which accounted for indirect costs reported savings of CAN$ 129 per patient. The certainty of evidence, using the GRADE assessment, for each outcome was very low.
Conclusion:
We found no evidence of additional adverse safety outcomes for once reprocessed cardiac catheterisation or implantable cardiac SUMDs. However, our confidence that the same findings would be observed in future studies is very low. There was insufficient evidence to establish the cost-effectiveness or environmental impacts of reusing cardiac catheterisation or implantable SUMDs. High-quality randomised controlled trials, analyses of national device reprocessing surveillance systems, cost-effectiveness studies, and life cycle assessments are required in order to facilitate better comparison across devices and reprocessing contexts.
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Types of Biopharmaceutical Studies: Controlled and Non-Controlled Approaches
Non-controlled studies, commonly employed for initial exploration, lack a control group, rendering them susceptible to biases and external influences. In contrast,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...