Thermoacoustic range verification during conventional dose rate delivery by a synchrocyclotron
Yao Hao1, Tuyen Mai2, Winter Green1
1Department of Radiation Oncology, Washington University in St Louis, St Louis, Missouri, USA.
A novel wireless thermoacoustic range verification system achieved millimeter accuracy for proton therapy, offering a rapid and non-invasive method for pinpointing the Bragg peak. This technology minimizes range uncertainties, enhancing treatment precision.
Area of Science:
- Medical Physics
- Radiation Oncology
- Ultrasound Technology
Background:
- Proton therapy requires precise range verification to ensure effective treatment delivery.
- Minimizing range uncertainties is critical for maximizing therapeutic outcomes and minimizing off-target radiation.
- Thermoacoustic range verification presents a promising non-invasive method for accurately locating the Bragg peak.
Purpose of the Study:
- To verify a prototype wireless thermoacoustic range verification device.
- To assess the device's minimal disruption to clinical workflows in proton therapy.
- To evaluate the device's performance with a synchrocyclotron-based proton therapy system.
Main Methods:
- A wireless thermoacoustic range verifier with an onboard radiation detector captured emissions from a gelatin ultrasound phantom.
- Proton beams were delivered using a Mevion Hyperscan S250i system with varying energy layers and lateral separations.
- Finite element acoustic simulations, informed by dose maps from RayStation 12A, were used to model beam current and analyze range shifts.
Main Results:
- The wireless range verification system achieved millimeter accuracy, with errors as low as -0.1 ± 1.3 mm when averaging 16 thermoacoustic pulses.
- Best- and worst-case errors across all spots were within ±1.6 mm, demonstrating robust performance.
- Analysis of single-layer data also showed high precision, with errors averaging 0.0 ± 1.3 mm in the best case.
Conclusions:
- A wireless research device for thermoacoustic range verification can be rapidly set up in under 10 minutes.
- The device demonstrated millimeter-level range accuracy in an ultrasound quality assurance phantom.
- This technology holds potential for improving the precision and efficiency of proton therapy.