Improving Pressure Steam Sterilization Quality Through Healthcare Failure Mode and Effects Analysis: A Pre-Post
Yanrong Zhang1,2, Ruixue Hu1,2, Yanhua Chen1,2
1Department of Sterile Processing Nursing, West China Second University Hospital, Sichuan University, Chengdu, Sichuan, People's Republic of China.
Objective:
To evaluate the effectiveness of Healthcare Failure Mode and Effects Analysis (HFMEA) in reducing quality defects during pressure steam sterilization in the Central Sterile Supply Department (CSSD).
Methods:
The study followed a structured HFMEA framework: (1) A multidisciplinary team (n=7) with CSSD expertise was established to analyze sterilization workflows, including instrument scanning, sterilization verification, and post-sterilization cooling. (2) Process mapping and risk prioritization were conducted using a 4-level severity/occurrence matrix (adapted from Australian clinical risk criteria) to calculate Risk Priority Numbers (RPN=Severity×Occurrence). High-risk failure modes (RPN≥8 or severity=4) were identified, including unlabeled "non-sterilized" packages (due to incomplete scanning), wet packages (from insufficient cooling<30 minutes), and unverified sterilization information. (3) Root causes were analyzed via fishbone diagrams (human, machine, material, environment, method). Targeted interventions included: optimizing the traceability system with department-specific alerts, standardizing scanning protocols, staff retraining on verification procedures, increasing instrument inventory and sterilizer racks, and implementing performance monitoring with 5W1H checklists.
Results:
Pre-intervention, 87 defects were identified among 185,382 sterilization packages (32 unlabeled "non-sterilized", 10 wet packages). Post-intervention, defects decreased to 11/189,531 packages (χ²=115.556, P<0.001), including 4 unlabeled (χ²=374.951, P<0.001) and 2 wet packages (χ²=8.889, P=0.003).
Conclusion:
Systematic HFMEA application reduced sterilization defects by addressing critical workflow gaps, demonstrating its value in enhancing CSSD quality control and patient safety.
More Related Videos
09:30Author Spotlight: Microbial Control and Monitoring Strategies for Cleanroom Environments and Cellular Therapies
Published on: March 17, 2023
10:03Investigating the Detrimental Effects of Low Pressure Plasma Sterilization on the Survival of Bacillus subtilis Spores Using Live Cell Microscopy
Published on: November 30, 2017
Related Concept Videos
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Healthcare Associated Infections II: Preventive Measures
The best practices for preventing healthcare-associated infections include hand hygiene, patient risk...
Errors occurring during blood pressure monitoring
Several factors...
Introduction to Statistical Process Control