Method Comparison and Investigation of Interference Variables of a Hand-Held Hemoglobinometer(HemoCue Hb 201+) in
Yuqing Sun1, Xiu Ting Yiew1, Janet Beeler-Marfisi2
1Departments of Clinical Studies, Ontario Veterinary College, University of Guelph, Guelph, Ontario, Canada.
Veterinary Clinical Pathology
|July 15, 2025
Summary
The HemoCue Hb 201+ (HC-201) provides reliable point-of-care hemoglobin measurements in cats, with minimal bias and excellent agreement compared to laboratory methods. This device aids in timely clinical decisions for anemic feline patients.
Area of Science:
- Veterinary Hematology
- Point-of-Care Diagnostics
- Clinical Pathology
Background:
- Monitoring anemic cats requires frequent blood sampling, risking further anemia.
- Laboratory hemoglobin results can cause delays in critical clinical decisions.
- The HemoCue Hb 201+ (HC-201) offers rapid, low-volume blood analysis for hemoglobin.
Purpose of the Study:
- Evaluate the diagnostic utility of the HC-201 for feline hemoglobin.
- Assess the impact of common blood interferents on HC-201 performance.
- Determine the reliability of HC-201 for point-of-care use in cats.
Main Methods:
- Compared HC-201 hemoglobin measurements to a reference analyzer (ADVIA 2120) using 154 samples from 93 cats.
- Assessed agreement and bias using Lin's concordance correlation coefficient, Bland-Altman, and Passing-Bablok regression.
- Evaluated HC-201 performance in the presence of anemia, leukocytosis, azotemia, lipemia, and different venipuncture sites.
Main Results:
- HC-201 showed excellent agreement (ρc=0.989) and precision (ICC=0.997) with the reference method.
- A small constant bias was detected, but no significant proportional bias was observed.
- Anemia, leukocytosis, azotemia, lipemia, and venipuncture site did not affect HC-201 results.
Conclusions:
- The HC-201 is a reliable point-of-care device for measuring hemoglobin concentration in cats.
- Its accuracy and precision support its use in clinical settings for rapid diagnostic assessment.
- This technology can improve efficiency in managing feline anemia and other conditions requiring hemoglobin monitoring.
Related Concept Videos
Equipments Used To Measure Blood Pressure
2.4K
Direct Method
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
2.4K
Atomic Absorption Spectroscopy: Interference
1.1K
Interference leads to systematic error in atomic absorption (AA) measurements by enhancing or diminishing the analytical signal or the background. These interferences can be grouped into three main categories: spectral interference, chemical interference, and physical interference.
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
1.1K