Apnea testing for brain death confirmation in VV-ECMO patients with very low sweep flow: a case reports and practical
Carine Carrijo de Faria1, Pedro Vitale Mendes1, Luis Carlos Cardoso Maia Junior1
1Intensive Care Unit, Hospital das Clínicas, Faculdade de Medicina, Universidade de São Paulo - São Paulo (SP), Brazil.
Abstract:
In recent years, venovenous extracorporeal membrane oxygenation has become a critical therapeutic tool for patients with severe respiratory failure. Neurological complications, including brain death, are common in this population, and confirming brain death in venovenous extracorporeal membrane oxygenation-supported patients presents unique challenges. In Brazil, an apnea test is mandatory for confirming brain death. However, its application in patients on venovenous extracorporeal membrane oxygenation, which predominantly addresses venoarterial extracorporeal membrane oxygenation cases, is not well defined in the literature. This report outlines our standardized approach for conducting apnea tests in three patients with suspected brain death during ongoing venovenous extracorporeal membrane oxygenation support. We describe three cases from a cohort of 93 extracorporeal membrane oxygenation patients treated for severe respiratory failure. The apnea test was conducted after 24 hours of observation without sedation. Given the physiological nuances of extracorporeal membrane oxygenation, where carbon dioxide clearance is primarily influenced by sweep flow, we adopted a low-sweep-flow protocol (200mL/minute) to achieve a partial pressure of carbon dioxide greater than 55mmHg, consistent with brain death criteria. In cases of severe hypoxemia during the test, extracorporeal membrane oxygenation blood flow can be temporarily increased to maintain oxygenation. All patients received concurrent renal support, which also facilitated carbon dioxide clearance. Our findings suggest that the apnea test with very low sweep flow is a safe and feasible method for diagnosing brain death in venovenous extracorporeal membrane oxygenation-supported patients. This physiologically grounded approach provides a clinically viable strategy for managing the complex interplay between gas exchange, oxygenation, and carbon dioxide clearance during the apnea test.
Related Concept Videos
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Cardiopulmonary Resuscitation II: ACLS Airway Management
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement: