Related Experiment Video
Updated: Sep 15, 2025

Roughness Impact of Piezoelectric Dental Scaler on Two Distinct Flowable Composite Filling Materials
Published on: January 10, 2025
[Comparative Study of Surface Conformity of Flexible Boluses Composed of Different Materials]
Hideharu Miura1,2, Kenji Kanemoto1, Toshiya Okazue1
1Hiroshima High-Precision Radiotherapy Cancer Center.
Purpose:
The purpose of this study was to compare the surface conformity of Clearfit Bolus II (Clearfit2) (Fujidenolo, Aichi) and Clearfit Bolus (Clearfit1) (Fujidenolo) using an anthropomorphic phantom.
Methods:
Clearfit1 and Clearfit2 boluses (5 mm thick) were placed in three anatomical regions of the anthropomorphic phantom: nose, breast, and abdomen, and CT scans were performed. Digital Imaging and Communications in Medicine (DICOM) structural data were used to measure the air gap between the bolus and the surface of the phantom. The surface conformity area ratio at the threshold (1-5 mm) was calculated. Statistical analysis was performed using the Mann-Whitney U test.
Results:
Clearfit2 had significantly reduced air gaps compared to Clearfit1: 2.9±2.0 mm in the nose, 1.8±2.1 mm in the breasts, and 0.1±0.1 mm in the abdomen, which were lower than Clearfit1's 5.8±2.9 mm, 11.4±8.4 mm, and 2.2±2.1 mm, respectively (p<0.001). In the abdominal region, Clearfit2 achieved near-perfect surface conformity with over 99% of the surface showing gaps ≤1 mm.
Conclusion:
We demonstrated that Clearfit2 significantly improves the surface conformity by reducing air gaps between the bolus and the phantom surface compared to Clearfit1.
More Related Videos
03:02Impact of Fabrication Techniques and Polishing Procedures on Surface Roughness of Denture Base Resins
Published on: January 17, 2025
12:07Three-dimensional Biomimetic Technology: Novel Biorubber Creates Defined Micro- and Macro-scale Architectures in Collagen Hydrogels
Published on: February 12, 2016
Related Concept Videos
One-Compartment Open Model for IV Bolus Administration: General Considerations
The drug's presence in the body is defined by an equation representing the difference between the rates of drug entry and exit. Key parameters—elimination rate constant,...
One-Compartment Open Model for IV Bolus Administration: Estimation of Clearance
In the one-compartment open model for intravenous (IV) bolus administration, clearance is estimated by dividing the elimination rate by the plasma drug concentration. This equation leverages the elimination rate constant and the apparent...
Two-Compartment Open Model: IV Bolus Administration
The disparity between drug input and the sum of drug transfer rates between...