Related Experiment Video
Updated: Sep 15, 2025

A Magnetic Resonance Imaging Protocol for Stroke Onset Time Estimation in Permanent Cerebral Ischemia
Published on: September 16, 2017
Automatic selection of optimal TI for flow-independent dark-blood delayed-enhancement MRI
Andreea Bianca Popescu1,2, Wolfgang Rehwald3,4, David Wendell4
1Department of Automation and Information Technology, Transilvania University of Brasov, Brasov, Romania.
Purpose:
Propose and evaluate an automatic approach for predicting the optimal inversion time (TI) for dark and gray blood images for flow-independent dark-blood delayed-enhancement (FIDDLE) acquisition based on free-breathing FIDDLE TI-scout images.
Methods:
In 267 patients, the TI-scout sequence acquired single-shot magnetization-prepared and associated reference images (without preparation) on a 3 T Magnetom Vida and a 1.5 T Magnetom Sola scanner. Data were reconstructed into phase-corrected TI-scout images typically covering TIs from 140 to 440 ms (20 ms increment). A deep learning network was trained to segment the myocardium and blood pool in reference images. These segmentation masks were transferred to the TI-scout images to derive intensity features of myocardium and blood, with which T1-recovery curves were determined by logarithmic fitting. The optimal TI for dark and gray blood images were derived as linear functions of the TI in which both T1-curves cross. This TI-prediction pipeline was evaluated in 64 clinical subjects.
Results:
The pipeline predicted optimal TIs with an average error less than 10 ms compared to manually annotated optimal TIs.
Conclusion:
The presented approach reliably and automatically predicted optimal TI for dark and gray blood FIDDLE acquisition, with an average error less than the TI increment of the FIDDLE TI-scout sequence.