Strut fracture in a Bjork-Shiley aortic valve prosthesis
Journal of Forensic Sciences
|October 1, 1985
Summary
Strut fracture in mechanical heart valves is a fatal risk. A Bjork-Shiley valve fractured, leading to a patient
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Pathology
Background:
- Mechanical prosthetic heart valves are used to replace damaged native valves.
- Strut fracture is a known, life-threatening complication associated with certain mechanical heart valve models.
- Previous incidents involved DeBakey, Beall, and Cooley-Cutter valves.
Observation:
- A 35-year-old male patient experienced sudden death 16 months post-implantation.
- The patient had a 60-degree Bjork-Shiley Convexo-Concave heart valve prosthesis in the aortic position.
- Autopsy revealed fractured welded attachments of the valve's outlet strut.
Findings:
- The fractured outlet strut attachments of the Bjork-Shiley valve were identified as the cause of failure.
- The valve's tilting disc was discovered dislodged in the patient's abdominal aorta.
- This case highlights a fatal strut fracture in the Bjork-Shiley Convexo-Concave model.
Implications:
- Strut fracture in mechanical heart valves poses a significant risk of mortality.
- This case underscores the critical need for ongoing surveillance and potential design improvements for prosthetic heart valves.
- Further research into the long-term durability and failure mechanisms of mechanical heart valves is warranted.