A Plasma Sample Preparation for Mass Spectrometry using an Automated Workstation
Published on: April 24, 2020
The Application of Six Sigma to Assess the Analytical Performance of Plasma Proteins and Design a Risk-Based
Ming Hu1, Jiaping Wang1, Huan Yang2
1Department of Laboratory Medicine, People's Hospital of Donghai County, Lianyungang, People's Republic of China.
Background:
This study applied the six sigma model to evaluate plasma protein testing performance in six laboratories, with customized quality control programs and targeted improvements introduced where necessary.
Methods:
Internal quality control (IQC) and external quality assessment (EQA) data for plasma proteins were gathered from six laboratories. Sigma values for each analyte were determined based on the coefficient of variation (CV), bias, and total allowable error (TEa). Using six sigma performance verification charts, we calibrated analyte performance and, guided by Westgard sigma rules, batch length, and quality goal index (QGI), developed laboratory-specific quality control schemes and improvement plans.
Results:
Despite standardized platforms and reagents, sigma values showed significant inter-laboratory variation, with some differences also observed within labs at varying analyte concentrations. For projects with sigma < 6, tailored quality control measures were implemented, leading to marked performance improvements.
Conclusion:
The six sigma model provides an objective framework for evaluating plasma protein test performance and enhancing quality. It enables quantitative assessment of laboratory management and supports the development and implementation of customized, risk-based statistical quality control (SQC) strategies and improvement measures across multiple laboratory systems.
More Related Videos
Related Concept Videos
Introduction to Statistical Process Control
Data Validation
Key parameters for method validation include: