Related Experiment Video
Updated: Sep 14, 2025

Reduced-gravity Environment Hardware Demonstrations of a Prototype Miniaturized Flow Cytometer and Companion Microfluidic Mixing Technology
Published on: November 13, 2014
Performance of Fluid Infusion Systems in the Changing Atmospheric Pressures Encountered in Aviation
Introduction:
With the increasing use of aeromedical transport for critically ill patients, it is essential to understand the impact of pressure changes on drug infusion delivery systems. As airplanes ascend and descend, gases/bubbles are released from solutions when ambient pressure decreases and dissolves when pressure increases. This may affect mechanical fluid delivery systems and cause clinically significant changes, especially within a critical care setting. We aimed to evaluate the impact of pressure changes on volumetric pumps and syringe drivers.
Methods:
An in vitro study of six volumetric pumps and eight syringe drivers was conducted in a hypobaric chamber to mimic pressure changes during flights. Infusion devices were set to deliver water at 0.2 ml ⋅ h-1 and infused volumes were measured. There were 15 open-ended syringes also studied.
Results:
During ascent, syringe drivers and volumetric pumps over-delivered 173 µL and 38 µL of fluid. During descent, syringe drivers under-delivered by 166 µL, whereas volumetric pumps under-delivered by 9 µL. Syringe drivers experienced statistically significant changes in fluid delivery during both ascent and descent. In volumetric pumps, only the descent phase infusion differed significantly from other phases. The volume of fluid expansion is dependent on volume and the mechanical properties of the fluid.
Discussion:
Decreasing ambient pressure causes bubble formation, which displaces fluid, and increasing ambient pressure causes bubble reabsorption in mechanical infusion devices. Hence, atmospheric pressure changes during air travel may alter fluid delivery from medical fluid delivery systems and affect critically ill patients who require both aeromedical evacuation and accurate infusion of drugs. Fan KS, Paterson M, Shojaee-Moradie F, Manoli A, Edwards V, Lee V, Hutchison E, Gifford RM, Parsons IT, Koehler G, Mathieu C, Mader JK, King BR, Russell-Jones D; EASA Consortium. Performance of fluid infusion systems in the changing atmospheric pressures encountered in aviation. Aerosp Med Hum Perform. 2025; 96(1):4-11.
More Related Videos
Related Concept Videos
Pressure Variation in a Fluid at Rest
When measuring pressure at two different levels within the fluid, the difference in...
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...
Pressure of Fluids
Vapor Pressure of Fluid
When a liquid is placed in a closed container with a small air space, and the space is evacuated, vapor molecules will...
Accelerating Fluids
The motion of the liquid within this infinitesimal cylinder is considered to obtain the pressure difference. Three vertical forces act on this liquid:
Bernoulli's Principle: Applications
Entrainment devices use a high fluid speed to create low pressures and, thus, entrain one fluid into another. Some examples of these devices are given below: