Optimizing fluorescence-based caries detection: A diagnostic performance analysis of SmarTooth and DIAGNOdent pen
Kyusik Kim1, Hoi-In Jung2, Wonse Park3
1Department of Pediatric Dentistry, Yonsei University College of Dentistry, Seoul, South Korea.
Objectives:
This study aimed to evaluate the diagnostic performance of DIAGNOdent (laser fluorescence [LF] pen; Kavo, Biberach, Germany), SmarTooth (SM; Smartooth, Korea), and the International Caries Detection and Assessment System II (ICDAS II) scoring system in detecting occlusal dental caries in permanent teeth, with micro-computed tomography (CT) as the gold standard. Furthermore, the optimal cut-off values for the two devices were determined.
Methods:
In total, 173 occlusal sites from extracted permanent teeth were assessed using the LF pen, SM, and ICDAS II. Radiographic values obtained using micro-CT were set as the gold standard. Sensitivity, specificity, accuracy, and optimal cut-off points were analyzed using receiver operating characteristic curve analysis. Significance was set at p < 0.05.
Results:
At the enamel demineralization (D1) and dentin demineralization (D2) thresholds, ICDAS II exhibited the highest sensitivity (D1: 0.966, D2: 0.897), whereas the LF pen exhibited the highest specificity (D1: 1.000, D2: 0.913). SM exhibited higher accuracy than did ICDAS II and the LF pen for both the D1 and D2 thresholds. The optimal cut-offs for the LF pen were as follows: sound: 0-9, enamel caries: 10-20, and dentin caries: 21-99; those for SM were as follows: sound: 0-8, enamel caries: 9-18, and dentin caries: 19-99.
Conclusions:
Both the LF pen and SM have high diagnostic performance for occlusal dental caries. However, SM achieves the best overall accuracy and thus has strong potential as an effective tool for caries detection.
Related Concept Videos
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a result, EDTA...
Tooth Anatomy
The Crown, Neck, and Root
The visible part of the tooth is referred to as the crown. It's covered by enamel, the hardest substance in the human body. The crown is uniquely shaped for each type of tooth, allowing for different functions such as cutting, tearing, or grinding food.
Assessment of the Mouth
Mouth Inspection
The inspection begins with visually examining the mouth for symmetry, color, and size.
Automated Microbial Diagnostics