In-vitro effects of modern radiotherapy regimes on cardiac implanted electrical devices
Steven Manley1,2, Amy Fenwick3, Megan Fraser3
1Department of Medical Physics, The James Cook University Hospital, Middlesbrough TS4 3BW, UK.
Background And Purpose:
External Beam Radiotherapy (EBRT) of patients with Cardiac Implanted Electrical Devices (CIEDs) have guidelines developed over many decades, during which both technologies have advanced. Consequently, guidelines may not reflect modern device interactions. Asynchronous modes, with pace sensing and shocks turned off whilst regulating pacing output, is used routinely for MRI scanning and could reduce risks for pace sensing errors in radiotherapy. Evidence is limited on modern radiotherapy using high dose, high doserate beams, with CIEDs in asynchronous mode. We present the effects of irradiating modern CIEDs using contemporary radiotherapy regimes.
Materials And Methods:
One hundred and sixty explanted CIEDs were irradiated, to corroborate historical findings for modern devices in normal operational modes, and explore effects when in asynchronous mode. Regimes knowingly detrimental to CIEDs; 48 Gy single fraction, neutron producing, alongside clinically relevant regimes of 60 Gy in 5 fractions using 10 MV flattening filter free [FFF] beams at clinical, and maximal dose rates.
Results:
No significant changes occurred to pacing voltages from 60 Gy in 5 fractions 10 MV FFF deliveries in asynchronous mode.No evidence supported restricting 6 MV flattened beams for CIED patients, including defibrillation capable devices.
Conclusions:
This study demonstrates asynchronous mode can reduce the frequency of CIED malfunctions during EBRT. However, clinical context, risks and benefit must be evaluated per patient. While some current guidelines potentially compromise plan quality to reduce dose to the CIED, the use of asynchronous mode may provide planning options, which more closely align to non-CIED cases.
More Related Videos
08:40Bidirectional Electrical and Optoelectronic Interfaces in Healthy and Ischemic Ex Vivo Rat Hearts
Published on: July 18, 2025
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Heart Failure Drugs: Inotropic Agents