An Improved Mechanical Testing Method to Assess Bone-implant Anchorage
Published on: February 10, 2014
The Utility of Evaluating Primary Stability of Implants Using Accumulated Insertion Torque Values: a Preliminary
Reiko Kobatake1, Kazuya Doi1, Tomoko Izumikawa1
1Department of Advanced Prosthodontics, Hiroshima University Graduate School of Biomedical and Health Sciences, HiroshimaJapan.
Objectives:
This preliminary in vitro study aimed to clarify the usefulness of the accumulated torque value as a new method for evaluating primary implant stability.
Material And Methods:
Implants of different sizes (42-10ST and 37-6ST) and simulated bone blocks of different densities (D1 to D4) were used. The implant osteotomy was prepared following the manufacturer's drilling protocol in simulated bone blocks, and the implants were inserted (n = 10). The implant insertion torque value was measured at 0.05 seconds intervals. The peak value was recorded as the peak torque value (PTV), and the sum of the torque values measured from the start to the end of the implant insertion was recorded as the accumulated torque value (ATV). After implantation, the implant stability quotient (ISQ) was measured. Correlations among ATV, PTV, and ISQ were analysed.
Results:
The ATV increased with the density of the simulated bone block. A strong correlation was observed between the PTV value and ATV value (42-10ST [P < 0.05, r = 0.99], 37-6ST [P < 0.05, r = 0.99]). In addition, a correlation was observed between the ATV and ISQ ISQ (42-10ST [P < 0.05, r = 0.81], 37-6ST [P < 0.05, r = 0.83]). The PTV was higher in the 42-10ST group than in the 37-6ST group from D1 to D4.
Conclusions:
The accumulated torque value varied according to the density of the simulated bone block, and a correlation was observed with the existing implant stability evaluation method. This suggests that this method may be useful as a novel approach for evaluating primary implant stability.
More Related Videos
07:32Author Spotlight: 3D Movement Assessment of Maxillary Posterior Teeth in Clear Aligner Treatment
Published on: February 23, 2024
09:07An Anesthesia, Surgery, and Harvest Method for the Evaluation of Transpedicular Screws Using an In Vivo Porcine Lumbar Spine Model
Published on: May 31, 2017
Related Concept Videos
Residual Stresses in Circular Shafts
Stability of structures
Screw: Problem Solving
To evaluate the...
Stress Concentrations in Circular Shafts