Optimized acquisition for simultaneous intravoxel incoherent motion and R2 quantification in the liver
Gregory Simchick1, Diego Hernando1,2
1Radiology, University of Wisconsin-Madison, Madison, Wisconsin, USA.
Purpose:
Design a b-value, first-order motion moment (M1), and echo time (TE)-optimized data acquisition for simultaneous intravoxel incoherent motion (IVIM) and R2 quantification in the liver.
Methods:
The b-M1-optimized (i.e., single TE) and b-M1-TE-optimized (i.e., multi-TE) IVIM acquisitions were determined using Cramer-Rao lower-bound optimization. Eleven healthy volunteers underwent MRI of the liver where each optimized acquisition was acquired twice (test-retest). For each data set, IVIM estimates (diffusion coefficient [D], perfusion signal fractions [f1 and f2], blood velocity standard deviations [SDs] [Vb1 and Vb2], and/or relaxation rates of tissue and blood [R2t and R2b]) were obtained in the right and left liver lobes within distinct spatial locations associated with slow (f1 and Vb1) and fast (f2 and Vb2) microvascular flow. Six R2-correction methods using combinations of estimated and/or fixed values of R2t and R2b were performed. Test-retest repeatability and interlobar agreement were evaluated using coefficients of variation (CVs) and relative mean differences (rMDs).
Results:
The b-M1-TE-optimized acquisition demonstrated acceptable to excellent right-lobe test-retest repeatability (CV = 4.58%, 8.94%, 8.62%, 27.5%, 18.8%, 2.85%, and 17.7% for D, Vb1, Vb2, f1, f2, R2t, and R2b, respectively) and interlobar agreement (|rMD| = 8.82%, 1.84%, 8.72%, 3.08%, and 0.20% for D, Vb1, Vb2, R2t, and R2b, respectively), except for interlobar agreement of f1 and f2 (rMD ≥ 37.7%). For the b-M1-optimized acquisition, R2 correction of f1 and f2 provided similar estimates as the b-M1-TE-optimized acquisition (|MD| ≤ 1.98%).
Conclusion:
The b-M1-TE-optimized IVIM acquisition enabled simultaneous and repeatable IVIM and R2 quantification in the liver. For single-TE IVIM data, R2 correction of f estimates using fixed values of R2t and R2b may be acceptable.