Spectral fluence response and off-axis corrections for ionization chambers and silicon diodes in small photon fields
Julian Roers1, Florian Katsch1, Hans T Eich1
1Department of Radiotherapy and Radiation Oncology, University Hospital Muenster, Albert-Schweitzer-Campus 1, Muenster, 48149, Germany.
Purpose:
This study investigates the charged particle fluence response of two reference ionization chambers (PTW T31021 Semiflex 3D and IBA CC04) and two silicon diode detectors (PTW microSilicon and IBA Razor diode) when measuring relative off-axis profiles in medium- to small-sized photon fields. Monte Carlo (MC) simulations are employed to determine the spectral fluence distribution of charged particles at different off-axis positions in water and within the investigated detectors. The goal is to enhance our understanding of the challenges associated with the dosimetry under non-equilibrium conditions and to calculate off-axis correction factors for the detectors under investigation.
Methods:
Spectral fluence correction factors pE,glf were determined using a modified version of the EGSnrc user code egs_chamber. A standard linear accelerator model based on the Elekta Precise was used as beam source with a photon beam quality specifier TPR20,10 of 0.671. Field sizes as small as 0.5 cm ×0.5 cm were evaluated at a water depth of 10 cm.
Results:
The analysis of spectral electron fluence in water as a function of off-axis positions across different field sizes revealed notable spectrum variations, particularly beyond the field edge and in the penumbra. These fluctuations directly impact the energy-dependent fluence response of the detectors. The assessment of pE,glf at the field edge demonstrated that the volume-averaging effect in air-filled ionization chambers is energy-dependent, with high-energy electron fluences being more strongly disturbed than low-energy fluences. In contrast, silicon diodes reproduced the relative profile within a 2% deviation but exhibited sensitivity to low-energy fluence fluctuations, especially in the penumbra.
Conclusion:
The findings highlight the importance of selecting appropriate detectors for relative profile measurements and the necessity of applying detector-specific off-axis correction factors to account for variations in spectral electron fluence response.
More Related Videos
10:16X-ray Beam Induced Current Measurements for Multi-Modal X-ray Microscopy of Solar Cells
Published on: August 20, 2019
06:49In situ Grazing Incidence Small Angle X-ray Scattering on Roll-To-Roll Coating of Organic Solar Cells with Laboratory X-ray Instrumentation
Published on: March 2, 2021
Related Concept Videos
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....
Atomic Emission Spectroscopy: Instrumentation
Atomic Emission Spectroscopy: Lab
Atomic Fluorescence Spectroscopy