Impact of patient transfer system on applicator stability and dosimetry in cervical cancer brachytherapy
Jittima Junhong1, Ekkasit Tharavichitkul2, Anirut Watcharawipha2
1Medical Physics Program, Department of Radiology, Faculty of Medicine, Chiang Mai University, Chiang Mai, Thailand.
Purpose:
This study aimed to evaluate applicator displacement due to patient transfer using traditional fixation (TRD) with and without patient transfer system (TS), and to assess its dosimetric impact on organs at risk (OARs).
Material And Methods:
Computed tomography (CT) images, including high-risk clinical target volume (HR-CTV) and OARs structures, were analyzed from 46 treatment plans involving 32 patients undergoing intracavitary brachytherapy with tandem-ovoid applicator, using either traditional fixation (TRD) alone or in combination with patient transfer system (TRD + TS). Applicator displacement was assessed by measuring the distance between applicator base in two sets of CT images: one acquired during treatment planning (pre-plan), and second at dose delivery (pre-load). Dosimetric impact was evaluated for each simulated applicator shift.
Results:
The use of TRD + TS resulted in reduced applicator displacement compared with TRD alone. The mean magnitude of displacement decreased from 2.94 mm to 1.43 mm laterally, from 3.17 mm to 1.83 mm in the anterior-posterior direction, and from 3.45 mm to 2.00 mm longitudinally. Regarding dose variations to D2cc for the bladder, rectum, sigmoid, and bowel, the average dose increases were 0.54%, 8.39%, 3.78%, and 0.58% for TRD, and 3.19%, 1.66%, 2.83%, and 1.19% for TRD + TS, respectively. No statistically significant difference was observed in either applicator displacement or radiation dose to OAR between the two systems.
Conclusions:
The combination of traditional fixation and patient transfer system have the potential to reduce applicator displacement, thereby enhancing treatment accuracy. However, in this study, no statistically significant difference was observed when compared with our traditional fixation method.
Related Concept Videos
Renal Failure: Dose Adjustments
Reduced renal clearance and elimination rate are common outcomes of renal impairment. These alterations lead to a prolonged elimination half-life and an altered apparent volume of distribution for drugs. As a result, dosage adjustments are typically necessary to maintain optimal drug levels in the body.
However, dosage adjustments...
Effect of Hepatic Disease on Pharmacokinetics: Drug Dosing and Hepatic Blood Flow
Effect of Hepatic Disease on Pharmacokinetics: Pathophysiologic Assessment and Liver Function Test
Drug Dosing in Renal Diseases: Dose Adjustments Based on Drug Clearance and Elimination Rate Constant
Kidney Transplant I: Introduction
Kidney Transplant III: Nursing Management