Related Experiment Video
Updated: Sep 14, 2025

Fully Automated Centrifugal Microfluidic Device for Ultrasensitive Protein Detection from Whole Blood
Published on: April 16, 2016
Ultra-low limit of detection of a carcinoembryonic antigen fiber sensor based on microfiber
Jinlong Wan1, Senyu Wang1, Sirong Zhou2
1State Key Laboratory of Electronic Thin Films and Integrated Devices, School of Optoelectronic Science and Engineering, University of Electronic Science and Technology of China (UESTC), Chengdu 610054, China.
Abstract:
Fiber optic sensors have been widely applied to address issues such as cardiovascular diseases, pathogen detection, and cancer biomarker detection. In this work, an ultra-low limit of detection, high-specificity, and label-free optical fiber biosensor for the detection of carcinoembryonic antigen (CEA) was proposed and experimentally demonstrated. A microfiber interferometer with refractive index sensitivity of up to 1915.955 nm/RIU was employed as the sensor head. The sensitivity of the sensor was further enhanced by the incorporation of molybdenum disulfide (MoS2), which facilitated the immobilization of CEA antibodies as specific detection markers on the microfiber surface. Experimental results showed that the proposed sensor achieved a detection limit of 15.86 fg/mL. The sensitivity of the biosensor was 4.727 nm/log(mg/mL). Notably, this biosensor demonstrated excellent specificity and label-free operation and enabled the detection of low-concentration CEA in approximately 15 minutes. The proposed micro fiber CEA sensor provided a highly sensitive and low-cost method for the early diagnosis of cancer.