A modular deep learning pipeline for enhanced plane-wave beamforming and B-mode image quality
Hamza Hadri1, Abderahhim Fail1, Mohamed Sadik1
11NEST Research Group, LRI Lab, ENSEM of Hassan II University of Casablanca., Casablanca, Morocco.
Background:
In ultrasound imaging using plane-wave (PW) techniques, image quality and contrast often suffer, especially when examining anechoic structures. Traditional beamforming methods like Delay-and-Sum or coherent PW compounding face limitations in balancing resolution and frame rate, which can result in suboptimal diagnostic accuracy.
Purpose:
This study aims to introduce a modular beamforming pipeline that overcomes these challenges and enhances PW image quality. By dividing the beamforming process into two modules: a multi-attention U-Net based model for capturing complex dependencies in time-delayed data and a super-resolution model for scaling up to the original B-mode image grid. We seek to improve PW image quality and modularity in the ultrasound imaging process.
Methods:
We implemented a modular beamforming approach, comprising a multi-attention U-Net model and a super-resolution model. We conducted experiments using simulated, experimental, and in vivo data from the PICMUS dataset to evaluate the performance of our pipeline against conventional methods such as PW1, PW9, and U-Net. Key metrics assessed included contrast-to-noise ratio (CNR), contrast ratio (CR), generalized Contrast-to-Noise Ratio (gCNR), and resolution.
Results:
Our model demonstrated superior performance across all metrics. In simulated data, the model achieved a 0.99 improvement in CNR, a 3.5 dB increase in CR, and 18% in gCNR compared to PW1. Experimental data showed a 0.7 enhancement in CNR and a 8.6 dB improvement in CR and 24% increase in gCNR when compared to PW1. In vivo data also revealed significant improvements, with a 1.9 dB increase in CR and a 0.15 enhancement in CNR over PW1. The enhanced performance in the anechoic cyst region underscores the model's effectiveness in improving image quality.
Conclusions:
The proposed modular beamforming approach offers significant advantages, including adaptability and improved image quality, despite the complexity of managing two models concurrently. The pipeline's flexibility in frame rate and image quality allows for customization based on specific clinical applications, making it a promising alternative to traditional methods.