Intra and inter-system reliability of a markerless motion capture system in different configurations during
Mark W Lenhoff1, Silvia Zanini1, Gregory Chingas1
1Department of Rehabilitation, Hospital for Special Surgery, 535 E 70th ST, NY 10021, USA.
Background:
Markerless motion capture is faster than a traditional marker-based method, but its reliability is not well established. The limitations of this new technology must be understood to ensure its appropriate utilization.
Research Question:
Does a markerless motion capture system exhibit intra-device and inter-device reliability with an Intraclass Correlation Coefficient (ICC) > 0.70 during functional tasks?
Methods:
Twenty test subjects were asked to squat and walk three times to compute the test-retest (repeated task with the same scaled skeletal model applied) and remove-replace (repeated task with two independently generated skeletal models applied) intra-device reliability for each of three devices and the inter-device reliability between the devices. Two devices were configured in a standard volume, and one device was configured sub-optimally in a larger volume.
Results:
For intra-device reliability, the average squat range of motion (ROM) test-retest ICC was 0.86 and average squat remove-replace ICC was 0.74. The average walk ROM test-retest ICC was 0.70 and the average walk remove-replace ICC was 0.72. For inter-device reliability between the two devices in the standard volume, the average ICC was 0.82 for squat and 0.72 for walk. With the sub-optimally configured device included for a three-device analysis the average inter-device reliability ICC dropped to 0.55 for squat and 0.46 for walk.
Significance:
Larger field of view, suboptimal camera placement and tasks performed away from the center of the field of view presented challenges to the DARI markerless mocap system. Intra-device and inter-device reliability can achieve a level of ICC > 0.7 using DARI markerless mocap technology using an optimized setup, but the transverse plane kinematics exhibit the weakest performance. Optimal camera placement (i.e., field of view) is critical for reliable results.