Non-invasive meningitis screening in neonates and infants: multicentre international study

Sara Ajanovic1,2,3,4, Beatrice Jobst5, Javier Jiménez5

  • 1ISGlobal, Barcelona, Spain. sara.ajanovic@isglobal.org.

Pediatric Research
|July 23, 2025
PubMed
Abstract

Insights

A new non-invasive ultrasound device accurately screens for meningitis in infants. This technology could reduce the need for lumbar punctures, especially in low-resource settings.

Area of Science:

  • Medical Imaging
  • Pediatric Infectious Diseases
  • Artificial Intelligence in Medicine

Background:

  • Meningitis diagnosis typically requires lumbar puncture (LP) for cerebrospinal fluid (CSF) analysis.
  • LPs are standard in high-income countries but rarely performed in low- and middle-income settings, leading to empirical treatment.
  • There is a need for non-invasive diagnostic tools for meningitis screening.

Purpose of the Study:

  • To validate a non-invasive transfontanellar white blood cell (WBC) counter in CSF for meningitis screening.
  • To assess the accuracy of a deep-learning algorithm using high-resolution ultrasound (HRUS) for CSF analysis.

Main Methods:

  • Prospective study across five hospitals (Spain, Mozambique, Morocco) from 2020-2023.
  • Included patients under 24 months with suspected meningitis and an open fontanelle.
  • Used HRUS to obtain CSF images, with a deep-learning model trained to classify WBC counts.

Main Results:

  • The algorithm analyzed 3782 images from 76 patients.
  • Achieved 94.4% sensitivity and 94.8% specificity in classifying CSF with 30 WBC/mm3.
  • One false negative occurred with a traumatic LP.

Conclusions:

  • The non-invasive ultrasound device is an accurate tool for screening meningitis in neonates and infants.
  • This technology can help modulate lumbar puncture indications, reducing procedures in low-suspicion cases.
  • It offers a vital alternative for critically ill patients and in resource-limited settings.