Validation of a popular consumer-grade cuffless blood pressure device for continuous 24 h monitoring
Bhavini J Bhatt1, Haashim Mohammad Amir1, Siana Jones1
1Institute of Cardiovascular Science, University College London, Gower Street, London WC1E 6BT, UK.
Insights
This study found that a popular wearable blood pressure monitor showed poor agreement with a medical-grade device, suggesting it may not replace standard 24-hour monitoring for hypertension detection.
Area of Science:
- Cardiology
- Biomedical Engineering
- Medical Devices
Background:
- Hypertension is a major global health concern, with many cases undiagnosed.
- Wearable, cuffless blood pressure (BP) monitors offer potential for widespread BP monitoring.
- The accuracy of consumer-grade wearable BP monitors requires rigorous validation.
Purpose of the Study:
- To validate a consumer-grade wearable BP monitor (W-BPM) against a medical-grade ambulatory BP monitor (A-BPM).
- To assess the accuracy of the W-BPM in identifying high blood pressure.
- To evaluate the W-BPM's ability to measure BP variability and nocturnal dipping.
Main Methods:
- Thirty-one participants simultaneously wore a W-BPM (Aktiia) and an A-BPM (Mobil-O-Graph) for 24 hours.
- Systolic BP (SBP), diastolic BP (DBP), and heart rate (HR) were recorded at set intervals and during rest.
- Agreement was assessed using standard statistical methods, including limits of agreement and correlation coefficients.
Main Results:
- The W-BPM showed poor to moderate agreement with the A-BPM (e.g., limits of agreement ~ -30/+30 mmHg for SBP).
- The W-BPM significantly underestimated BP variability and nocturnal BP dipping.
- Sensitivity and specificity for detecting high BP were approximately 50% and 80%, respectively.
Conclusions:
- The cuffless W-BPM demonstrated low agreement with cuff-based ambulatory monitoring.
- The device may not be a suitable replacement for current gold-standard 24-hour BP monitoring.
- Further research is needed to determine the clinical utility of cuffless BP monitors.
Aims:
Hypertension is a leading cause of death worldwide, yet many hypertensive cases remain undiagnosed. Wearable, cuffless blood pressure (BP) monitors could be deployed at scale, but their accuracy remains undetermined.
Methods And Results:
This study validated a popular consumer-grade wearable BP monitor (W-BPM, Aktiia), using a medical-grade ambulatory device (A-BPM, Mobil-O-Graph), for reference. Thirty-one participants (aged 19-62 years, 17 (55%) females, in office BP 121 ± 15 over 77 ± 12 mmHg) simultaneously wore both devices for 24 h. Systolic BP (SBP), diastolic BP (DBP), and heart rate (HR) were measured in pre-set intervals by the A-BPM and at rest by the W-BPM. Agreement was assessed using standard methods. Accuracy in identifying high BP (mean 24 h SBP/DBP > 130/80 mmHg) was assessed. Compared to A-BMP, mean SBP and DBP tended to be slightly lower during the day and not significantly different at night. Nocturnal BP dipping and BP variability were significantly underestimated by the W-BPM. Agreement between the two devices was poor to moderate (limits of agreement of about -30/+30 mmHg for SBP and -20/+15 mmHg for DBP, correlation coefficients between 0.20 and 0.42). Sensitivity and specificity for high BP detection were around 50% and 80%, respectively. Limiting the analysis to measures taken in similar conditions (within 10 min and with HR within ±10 b.p.m.) did not improve agreement.
Conclusion:
Low agreement suggests that the cuffless device may not be a suitable replacement for standard 24 h cuff-based ambulatory monitoring. Further data are required to assess the clinical role of cuffless BP monitors.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Holter Monitor: 24-Hour Monitoring
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...