Photoplethysmography in recent-onset atrial fibrillation: automatic detection of rhythm change and burden
Olli A Rantula1,2,3, Jukka A Lipponen4, Jari Halonen1,3
1School of Medicine, Faculty of Health Sciences, University of Eastern Finland, PO Box 1627, 70210 Kuopio, Finland.
Aims:
Atrial fibrillation (AF) is the most common arrhythmia, increasing stroke risk. Detecting AF is challenging due to its asymptomatic and paroxysmal nature. This study combines photoplethysmography (PPG) with automated techniques to detect AF, assess AF burden, and monitor rhythm changes from AF to sinus rhythm (SR).
Methods And Results:
Ninety patients with recent-onset (duration <48 h) AF, scheduled for cardioversion, were monitored using a three-channel PPG armband on the upper arm. An ambulatory three-lead electrocardiogram (ECG) served as the gold standard. PPG recordings were segmented into 10-, 20-, 30-, and 60-min detection windows. Automated detection identified SR and AF episodes, rhythm changes, and AF burden. Sensitivities, specificities, positive predictive values (PPVs), and negative predictive values (NPVs) for rhythm detection were calculated, and the intraclass correlation coefficients (ICCs) for PPG-based AF burden were compared to the gold standard. Monitoring time ranged from 1.0 to 8.2 h per patient. Sensitivities, specificities, PPVs, and NPVs for AF detection were 93.9-94.6, 99.5-99.8, 99.4-99.7, and 93.7-95.0%, respectively. The ICC (0.97-0.98) indicated excellent agreement between PPG and the gold standard in estimating AF burden, with differences of -6.3 to -8.3 min (5.5-6.8%). Rhythm changes from AF to SR were detected in all patients (sensitivity 100%), with detection delays of 4.1 ± 1.4, 8.7 ± 2.8, 13.7 ± 3.9, and 27.8 ± 7.1 min depending on the detection window.
Conclusion:
Photoplethysmography with automated analysis shows promise in detecting AF, AF burden, and rhythm changes, indicating its potential in AF screening.
Clinical Trial Registration:
NCT04917653.
More Related Videos
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Assessment of blood pressure in brachial artery(two-step method)
Dysrhythmias V: Evaluating Dysrhythmias
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...