The impact of sensor application site on continuous noninvasive haemoglobin trends
Audrius Andrijauskas1, Povilas Andrijauskas2, Darius Dilijonas3
1Clinic of Anaesthesiology and Intensive Care, Institute of Clinical Medicine, Faculty of Medicine, Vilnius University, M.K. Čiurlionio str. 21, Vilnius LT-03101, Lithuania.
Purpose:
This prospective observational clinical study investigated discrepancy between noninvasive continuous haemoglobin (SpHb, Masimo Corp., Irvine, CA) measurements with sensors placed on three fingertips during elective cardiac surgery.
Materials And Methods:
The study included 12 adult patients who underwent elective coronary artery bypass grafting (CABG) under cardiopulmonary bypass (CPB). Three identical versions of SpHb sensors were placed on the fingers of the same hand and connected to dedicated Masimo Radical-7 monitors. Data were continuously recorded throughout the surgery. Multiple group comparison tests-Kruskal-Wallis and Alexander-Govern tests-were used to compare three sets of SpHb data (trends). Bland-Altman and Giavarina plots were used to compare the SpHb data with invasive arterial haemoglobin (Lab-Hb) data obtained at clinically relevant time points chosen at the discretion of the cardiac anaesthesiologist, who was blinded to the study protocol.
Results:
There was a good correlation between the time-matched SpHb data from the three sensors at 253 time points (r = 0.86-0.95), but there was a significant difference between the SpHb values obtained from each sensor and the average SpHb value of all 3 sensors (p < 0.05) at each time point. The absolute difference between the SpHb value obtained from each sensor and the average SpHb value was 0.69 g/dl g/dl (SD 1.49 g/dl). The accuracy and precision of the average SpHb for estimating Lab-Hb was higher compared to the SpHb values obtained from two out of three individual sensors (sn1 and sn3) and very similar compared to values obtained from sensor sn2.
Conclusion:
The sensor placement site can impact the SpHb measurement causing discrepancy between SpHb measured in different fingers. The average SpHb from three fingertips appears to be more accurate and precise in estimating Lab-Hb. However, more advanced sensors are needed.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
05:18A Rapid and Chemical-free Hemoglobin Assay with Photothermal Angular Light Scattering
Published on: December 7, 2016
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...