COR-KNOT-induced aortic root injury during minimally invasive aortic valve replacement of a bicuspid aortic valve: a
Takahiro Ishigaki1, Kazuma Okamoto2, Satoshi Asada3
1Department of Cardiovascular Surgery, KKR Sapporo Medical Center, Hokkaido, Japan. t-ishigaki@kkr-smc.com.
Background:
The COR-KNOT device is widely used in minimally invasive cardiac surgery, and its efficacy in reducing operative time is well established. However, it has been associated with complications, such as valve leaflet perforation and prosthetic valve damage.
Case Presentation:
We report a rare case of aortic root injury caused by COR-KNOT usage. A 72-year-old man with severe aortic stenosis due to a bicuspid aortic valve and distorted aortic root anatomy underwent minimally invasive aortic valve replacement using the COR-KNOT. After aortotomy closure, vigorous bleeding from the aortic root was observed. Punctate injuries of the aortic root, corresponding to the COR-KNOT clip sites, were identified following re-aortic cross-clamping. The procedure was converted to median sternotomy, the aortic root injury was repaired with an autologous pericardial patch, and the bioprosthetic valve was re-implanted using hand-tied sutures. The patient recovered uneventfully.
Conclusions:
This case highlights the importance of careful device selection and clip positioning in anatomically challenging cases to avoid life-threatening complications.