Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Transseptal access through synthetic materials: An in vitro comparative study of radiofrequency and needle-based
Maxwell R Klein1, Dominic P Recco2, Nicholas E Kneier3
1Department of Cardiology, Boston Children's Hospital, Boston, Massachusetts.
Background:
The transseptal approach for accessing left heart structures is commonly performed with metal needles and radiofrequency (RF) needles and wires, though how each device performs when crossing materials used in congenital heart disease (CHD) surgery is unknown.
Objective:
To compare the efficacy of commercially available transseptal devices in puncturing select materials relevant to CHD surgery.
Methods:
A custom-designed in vitro transseptal model was created. Transseptal devices included a metal needle, an RF needle, and an RF wire. Study materials included expanded polytetrafluoroethylene (ePTFE), polyester, and bovine pericardium; porcine atrial septum was the control material. The primary outcomes were (1) the number of 1-second RF pulses required to puncture for RF devices and (2) the peak force required to puncture for the metal needle.
Results:
Excluding ePTFE, RF-based devices punctured all materials at 5-10 g force with a single 1-second pulse. For ePTFE, the RF wire required a median of 4 seconds (range 2-6 seconds) of RF energy to puncture, whereas the RF needle failed entirely. The metal needle required higher forces across all materials, most drastically for polyester (median 277 g). The metal needle punctured ePTFE with reasonable force (median 68 g).
Conclusion:
Commercially available RF transseptal platforms successfully puncture native septa, polyester, and bovine pericardium with short-duration RF energy delivered at low forces. Metal needles remain a cost-effective option when crossing native septum, ePTFE, and bovine pericardium. However, caution should be taken when puncturing polyester with the metal needle, given that substantially higher applied forces are required. The RF wire and metal needle are reasonable options for ePTFE puncture.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
07:39Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for diagnosing...