Related Experiment Video
Updated: Sep 13, 2025

Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Diagnostic performance of single-lead electrocardiograms from a smartwatch and a smartring for cardiac arrhythmia
Andre Briosa E Gala1,2,3, Alexander James Sharp1,4, David Schramm1
1Department of Cardiology, Oxford University Hospitals NHS Foundation Trust, Oxford, United Kingdom.
Background:
Wearable devices are widely used for atrial fibrillation (AF) detection, yet most validation studies include only sinus rhythm or AF, likely overestimating diagnostic performance.
Objective:
This multicenter study assessed the performance of automated AF detection and physician interpretation of single-lead electrocardiograms (SL-ECGs) from the Apple Watch and CART Ring.
Methodology:
Participants underwent simultaneous 12-lead ECG and SL-ECGs from Apple Watch and CART Ring. Two cardiologists independently adjudicated all ECGs. Apple Watch and CART Ring classified recordings as "AF," "Not AF," or "Unclassified." Diagnostic performance for automated AF detection was evaluated in "worst-case" (all SL-ECGs) and lenient (excluding unclassified SL-ECGs) scenarios. Physician interpretation of SL-ECGs was also compared to 12-lead ECG.
Results:
Among 483 patients (median age, 66 years; 29% female), 196 (39%) had AF across 3 United Kingdom centers. A total of 2398 ECGs were analyzed. Interobserver variability was excellent (Cohen's kappa: Apple Watch, 0.85; CART Ring, 0.84). In the "worst-case" analysis, CART Ring outperformed Apple Watch (sensitivity, 84.6% vs 69.1%; specificity, 89.9% vs 72.6%). Apple Watch had more unclassified SL-ECGs (20.1%) than CART Ring (1.9%). The lenient analysis showed an improvement in sensitivity (CART Ring, 84.8 %; Apple Watch, 86.4%) and specificity (CART Ring, 91.2%; Apple Watch, 91.7%). Physician interpretation improved diagnostic performance for AF and sinus rhythm but remained limited for other arrhythmias.
Conclusion:
Apple Watch missed approximately 1 in 3 episodes of AF and a high number of unclassified SL-ECG. CART Ring demonstrated superior performance. Physician interpretation significantly improved AF diagnosis but remained unreliable for other arrhythmias, emphasizing the need for cautious integration of wearable ECGs into clinical practice.
More Related Videos
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Holter Monitor: 24-Hour Monitoring
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Mechanism of Cardiac Arrhythmias