Foot Measurements From Three-Dimensional Scans: Elinvision 3DST Reliability
Petra J Jones1,2, Alex V Rowlands2,3,4, Melanie J Davies1,2,3
1Leicester Diabetes Centre, University Hospitals of Leicester, Leicester General Hospital, Leicester, UK.
The 3DST foot scanner shows excellent reliability for basic foot measurements, but accuracy limitations for certain dimensions hinder its use in custom orthotics. Further research is needed for therapeutic footwear applications.
Area of Science:
- Biomechanics
- Podiatry
- Medical Device Technology
Background:
- Traditional plaster casting for orthotics and therapeutic footwear is time-consuming.
- 3D foot scanners offer a potentially faster alternative for capturing foot dimensions.
Purpose of the Study:
- To evaluate the reliability of the Elinvision 3DST foot scanner for key foot measurements.
- To compare 3D scanner measurements with traditional manual measurements.
Main Methods:
- A repeated measures design was employed.
- Two raters performed three scans of the right foot for 20 healthy adults using the 3DST scanner.
- Manual measurements were taken by an experienced rater using a Ritz stick and tape measure.
Main Results:
- Excellent inter-rater and intra-rater reliability (Intraclass Correlation Coefficients 0.99-1.00) were observed for the 3DST scanner.
- Scanner and manual measurements showed high agreement for foot length, ball width, heel width, and ball girth (≤0.4 cm difference).
- Larger discrepancies were found for foot waist, short heel, ankle circumference, and anatomical ball width (0.5-1.1 cm difference).
Conclusions:
- The 3DST scanner demonstrates high reliability and potential for capturing basic foot dimensions in footwear fit research.
- Significant differences in specific measurements limit its current application in the production of custom orthotics and therapeutic footwear.