Flexural strength of translucent zirconia for single crowns and fixed dental prostheses-A systematic review
Selina A Bernauer1, Nicola M Lirgg1,2, Alexis Ioannidis3
1Department of Reconstructive Dentistry, University Center for Dental Medicine Basel UZB, University of Basel, Basel, Switzerland.
Purpose:
The aim of this systematic review was to analyze the reported flexural strength values of commercially available translucent zirconia materials. Additionally, data were evaluated to assess the impact of different specimen preparations and test methods.
Study Selection:
This systematic review was conducted according to the PRISMA guidelines. Electronic searches were performed to identify in vitro studies published between January 1, 2013, and March 15, 2025, that reported flexural strength values of translucent zirconia and included the names of the tested materials and manufacturers. Flexural strength values were analyzed to assess the effects of yttria content, specimen processing, surface treatments, and test methods using Kruskal-Wallis analysis of variance (α = 0.05) and forest plots.
Results:
A total of 78 laboratory studies met the inclusion criteria. The overall flexural strength values were significantly higher for 4Y-PSZ (803± 233 MPa) than for 5Y-PSZ materials (570 ± 116 MPa) (P = 0.005). Specimen processing, surface treatments, and test methods significantly affected the flexural strength values for 5Y-PSZ (P < 0.001) but not for 4Y-PSZ (P > 0.05).
Conclusions:
This systematic review highlights the challenges in comparing the flexural strength values of translucent zirconia owing to differences in processing parameters. Further standardization of test methods and adherence to ISO 6872 would improve comparability. Our findings support the clinical use of 4Y-PSZ for three-unit fixed dental prostheses (FDPs) and 5Y-PSZ for three-unit FDPs in the anterior region based on ISO 6872.
Related Concept Videos
Bonding and Strength of Aggregate
Flexural Stress
Hooke's Law states that within the material's elastic limits, stress is directly proportional to strain. In a member experiencing a bending moment, the strain at any point is relative to...
Tensile Strength Considerations of Concrete
The dimensions and shape of a concrete specimen...
Fiber Reinforced Concrete
Impact Strength of Concrete
Strength of Cement
For compressive strength tests, ASTM C 109-05 standards prescribe a cement-sand mix ratio of 1:2.75 and a water/cement ratio of 0.485 for making 2-inch cubes. These cubes are mixed, cast, and cured in saturated lime water at 23°C until testing. Flexural strength testing, outlined in...