Clinical note: the in-vivo demagnetisation and successful re-magnetisation of a cochlear implant magnet using 3T
Francisco Ropero Romero1, María Del Mar Barrios-Romero2, Marta Bastarrica Marti2
1Servicio de Otorrinolaringología, Hospital Universitario Virgen Macarena, Sevilla, Spain.
Objective:
This case report describes the successful in-vivo application of a procedure to re-magnetise the cochlear implant (CI) magnet in a unilaterally implanted user with an older generation implant, following an inadvertent exposure to a higher strength magnetic resonance imaging (MRI) field for which the implant was indicated.
Design:
A case report.
Study Sample:
An adult with a unilateral CI (a SONATA TI100 CI).
Results:
Re-magnetisation was accomplished using the static magnetic field of a 3.0 T MRI scanner. It should be noted that the described re-magnetisation procedure is appropriate only for 1) unilateral CI users and 2) older generation CIs which have a non-removable or a well-secured magnet which can withstand MRI scanners only up to 1.5 T.
Conclusions:
This is the first study to successfully demonstrate the partial re-magnetisation of a CI magnet using a 3.0 T MRI scanner in-vivo. This procedure is a potential option to resolve de-magnetisation issues without surgery or device replacement.
More Related Videos
11:27Magnetically-Assisted Remote Controlled Microcatheter Tip Deflection under Magnetic Resonance Imaging
Published on: April 4, 2013
05:26Author Spotlight: Low-Cost Electroencephalographic Recording System Combined with a Millimeter-Sized Coil to Transcranially Stimulate the Mouse Brain In Vivo
Published on: May 26, 2023
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI