Visualization of Intensity Levels to Reduce the Gap Between Self-Reported and Directly Measured Physical Activity
Published on: March 7, 2019
Physical activity cut-points for older adults using the Zio XT onboard accelerometer
Lacey H Etzkorn1,2, Anis Davoudi1, Erin E Dooley3
1Department of Epidemiology, Bloomberg School of Public Health, Johns Hopkins University, Baltimore, MD, USA.
New cut-points for the Zio XT monitor help estimate physical activity intensity. The Zio accelerometer provides reliable activity estimates comparable to waist-worn devices, aiding clinical research in older adults.
Area of Science:
- Biomedical Engineering
- Physical Activity Measurement
- Gerontology
Background:
- The Zio XT monitor, an electrocardiographic device, includes an accelerometer for quantifying physical activity.
- Accurate physical activity assessment is crucial for understanding health in free-living environments, especially in older adults.
Purpose of the Study:
- To establish activity cut-points for the Zio accelerometer to estimate time spent in very light, light, and moderate-to-vigorous physical activity (VLIPA, LIPA, MVPA).
- To compare Zio's 24-hour wear protocol estimates of daily MVPA with a waist-worn accelerometer's waking-wear protocol.
Main Methods:
- Participants (n=381) from the ARIC study simultaneously wore Zio and waist-worn ActiGraph GT3X devices.
- Physical activity was summarized using Mean Amplitude Deviation (MAD) for Zio and Vector Magnitude Counts (VMC) for GT3X.
- GT3X VMC cut-points were mapped to Zio MAD using a quantile-quantile approach.
Main Results:
- Mapped Zio MAD cut-points for VLIPA, LIPA, and MVPA were 9.04, 28.2, and 58.1 mili-g, respectively.
- Zio and GT3X showed strong correlations (ICC > 0.7) for daily activity intensity when wear time overlapped.
- MVPA estimates showed high agreement (ICC = 0.77) between devices using their respective wear times; Zio reported more LIPA and VLIPA.
Conclusions:
- Recommended cut-points for Zio accelerometry in clinical research involving older adults.
- Zio accelerometry provides reliable physical activity intensity data comparable to traditional waist-worn devices.
Related Concept Videos
Average Acceleration
Power Expended by a Constant Force
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Mechanical Efficiency of Real Machines
However, in reality, no machine can be truly ideal, and all of them experience some...
Motor Unit Stimulation
The latent period of contraction marks the onset of excitation-contraction coupling, when the action potential propagates across the sarcolemma, preparing the muscle fibers for contraction. As the fibers enter the contraction phase, the...
Muscle Stimulation Frequency
Wave summation
At low firing rates, motor neurons induce individual twitch contractions in muscle fibers. These twitches...