Measuring Blood Pressure in Mice using Volume Pressure Recording, a Tail-cuff Method
Published on: May 15, 2009
Evaluation of the arterial blood pressure of dogs by two noninvasive methods
Abstract:
The determination of the arterial blood pressure was done on 12 healthy mixed breed dogs in both the anesthetized and the conscious state, to evaluate two instruments (Doppler flow detector and infrasonde D4000), in their ability to indirectly determine arterial blood pressure. The coefficients of variation were higher with indirect methods when compared with the results obtained by cannulation. These coefficients were lower with the Doppler flow detector. The correlation study showed that both apparatuses were reliable in most situations. The infrasonde D4000 was more accurate than the Doppler in the conscious animals. However the results showed a lack of precision in hypertensive conscious dogs. The diastolic arterial blood pressure was particularly precise in the case of the anesthetized hypotensive dogs. Its sensitivity allowed it to register muscle movement artifacts. The Doppler flow detector showed less variation and was particularly accurate in both anesthetized and conscious hypertensive dogs. Its sensitivity allowed artifact movement sounds to be detected. The Doppler should be used in quiet surroundings or earphones should be worn by the evaluator. Some form of restraint is needed with the use of both instruments. Even if the correlations with the direct arterial blood pressure values were better with the infrasonde D4000, greater variations were found in the individual readings. The Doppler instrument represents in the hands of the investigators a better instrument for routine monitoring of blood pressure in the dog.
More Related Videos
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
07:35Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.