Monitoring Lung Function with Electrical Impedance Tomography in the Intensive Care Unit
Published on: September 6, 2024
Comparison of regional lung mechanics using electrical impedance tomography in mechanically ventilated COVID-19 vs
Eda Aydeniz1, Serge J H Heines2, Frank van Rosmalen1
1Department of Intensive Care Medicine, Maastricht University Medical Center+, Maastricht, the Netherlands; Cardiovascular Research Institute Maastricht (CARIM), Maastricht University, Maastricht, the Netherlands.
Background:
Whether coronavirus disease 2019 acute respiratory distress syndrome (COVID-ARDS) and acute respiratory distress syndrome (ARDS) have a different phenotype remains subject of research and debate. We aimed to study the differences and similarities of electrical impedance tomography (EIT) parameters in patients with and without COVID-19.
Methods:
This retrospective observational study included patients with an EIT measurement during ICU admission. EIT variables, including dynamic compliance, alveolar overdistension and collapse, were visualized for each positive end-expiratory pressure (PEEP) level of the EIT measurement using polynomial regression. Two-level linear mixed-effects regression with random intercept and slope for PEEP was performed for alveolar overdistension per PEEP level during EIT measurement in COVID-19 and non-COVID-19 patients.
Results:
From October 2013 until October 2023, 131 COVID-19 patients and 128 non-COVID-19 patients underwent an EIT measurement. The overall pattern of dynamic compliance, alveolar overdistension and collapse was similar in COVID-19 and non-COVID-19 patients. Visual inspection of EIT variables showed a higher mean alveolar overdistension in COVID-19 patients compared to the non-COVID-19 patients (day 1-3: maximum 43.2 % vs maximum 26.1 %, day 4-6: maximum 46.8 % vs maximum 25.2 %, ≥7 days: maximum 41.4 % vs maximum 33.0 %). However, two-level linear mixed-effects regression with random intercept and slope for PEEP showed no statistically significant difference between COVID-19 and non-COVID-19 patients (-1.0 % [-3.5; 1.6], p = 0.459).
Conclusions:
Regional lung dynamics were generally comparable in COVID-19 and non-COVID-19 patients when assessed by EIT. Based on these findings, mechanical ventilation protocols should be optimized for comparable parameters in COVID-19 and non-COVID-19 patients.
More Related Videos
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation I: Indication and Settings
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...