Dual flip angle liver T1 mapping with B1 correction at 3 T - In vitro and in vivo evaluation
Kim Sivesgaard1, Rozh Al-Mashhadi1, Thomas Winther Buus2
1Department of Radiology, Aarhus University Hospital, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark; Department of Clinical Medicine, Aarhus University, Palle Juul-Jensens Boulevard 99, 8200 Aarhus N, Denmark.
Purpose:
To investigate the in vitro precision and linearity of 3D whole-liver covering dual flip angle (DFA) T1 mapping and compare different circular ROI sampling strategies in vivo.
Methods:
All scans were performed at 3 T using a commercially available three-dimensional steady-state spoiled gradient echo with DFA. The in vitro study was performed on the EUROSPIN Test Object 5 with inversion recovery turbo spin echo (IRTSE) performed for reference T1 values. The in vivo study was performed in 30 patients before and after gadoxetic acid. In vitro DFA T1 map was compared to IRTSE T1 map voxel-by-voxel. In vivo DFA T1 maps were manually segmented for mean-segment T1. Two readers estimated segmental T1 using 4 ROI sampling strategies. Inter-reader agreement was assessed, and ROI sampling strategies were compared to manual segmentation. Linear regression, Bland-Altman plots and comparison of correlations were performed.
Results:
In vitro DFA T1 measurements showed strong linearity with R2 = 0.9912 (P < 0.0001) compared to IRTSE with 16.33 % overestimation. In vivo ROI-based estimated segmental T1 demonstrated improved inter-reader agreement with R2 increasing from 0.9135 to 0.9601 (P < 0.003) and narrower limits-of-agreement as the ROI area increased. Agreement with manual segmentation-based mean-segment T1 improved with increasing ROI area (R2 0.9718 to 0.9778 and 0.9359 to 0.9722, for reader 1 and 2 respectively) as well as narrower limits-of-agreement.
Conclusion:
DFA T1 measurements had a strong linearity with a minor overestimation. The use of larger ROI areas leads to better agreement with manual segmentation and improved inter-reader agreement.
More Related Videos
08:42Using Multi-fluorinated Bile Acids and In Vivo Magnetic Resonance Imaging to Measure Bile Acid Transport
Published on: November 27, 2016
08:41Novel In Vivo Micro-Computed Tomography Imaging Techniques for Assessing the Progression of Non-Alcoholic Fatty Liver Disease
Published on: March 24, 2023
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography