X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Determination of redundant scan coverages along the Z-axis and dose implications for common paediatric computed
Rosemary Yeboah Ampofo1, Kofi Adesi Kyei2,3,4, Joseph Daniels5
1Department of Radiography, University of Ghana, P.O.Box 143, Legon, Accra, Ghana.
Introduction:
Optimizing scan parameters, particularly along the Z-axis, is crucial for minimizing unnecessary medical radiation exposure. Understanding and mitigating redundant scan coverages is essential for enhancing the safety and efficacy of paediatric CT imaging. This study quantified redundant Z-axis scan coverage and evaluated its associated dose implications in paediatric CT examinations in a limited-resource healthcare setting.
Methods:
A retrospective review was conducted on 279 paediatric CT scan records from two large tertiary hospitals in Ghana. Distances above the upper target and below the lower target of selected anatomical regions were measured using calibrated callipers on the CT console. The National Cancer Institute Dosimetry System for Computed Tomography (Monte Carlo-based software) was used to simulate the scanning situations and organ-dose implications.
Results:
In all, 87.3% of the CT scans had redundancies ranging from 0.77 ± 0.42 to 2.34 ± 1.22 cm for head, 1.43 ± 2.36 to 5.64 ± 3.00 cm for chest and 4.86 ± 3.01 to 6.75 ± 4.28 cm for abdominopelvic CT scans. Optimizing the scan length to the appropriate anatomical boundaries reduced the total dose-length product of neonates and middle-aged children by 27.9% and 26.1% respectively for chest CT scans, and 26.2% for infants during abdominopelvic CT scans. Optimal scan length selection for chest CT examinations resulted in organ-dose reductions of 57.5%, 56.1%, 63.6% and 73.8% for the thyroid glands, heart wall, lungs and breasts respectively.
Conclusion:
The study demonstrates a high prevalence (87.3%) of redundant scan coverage in paediatric CT examinations, with significant variances across different body sites. Implementing optimized scan lengths could substantially reduce radiation exposure and significantly lower organ doses.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation I: X-ray and CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...