Evaluation of Concomitant Imaging Dose in 4D-CBCT Guided Thoracic Radiotherapy
Wu Jianan1, Sang Yong1, Li Zhangmin2
1Department of Radiation Oncology, National Cancer Center/National Clinical Research Center for Cancer/Cancer Hospital & Shenzhen Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Shenzhen, People's Republic of China.
None:
IntroductionThis retrospective study evaluated the imaging dose of 4D-CBCT in patients treated with thoracic radiotherapy.MethodsA geometric model of a 4D-CBCT system was built for Monte Carlo (MC) calculations. Percentage depth dose (PDD) and profiles of the system were measured and compared with MC calculations to verify the model. Eight lung cancer patients were retrospectively selected for 4D-CBCT imaging dose assessment. For each patient, the imaging dose was calculated using the CBCT log file, simulating CT and the MC model. Organ at risk (OAR) dose parameters from the imaging dose were calculated, and the effect of the imaging dose on OAR dose parameters when the imaging dose was added to the treatment dose was analyzed.ResultsDeviations between MC calculations and measurements for PDD and profiles were within 1.1% and 5%, respectively. For a pre-treatment 4D-CBCT scan, the mean lung and heart doses were 9.5 ± 1.4 mGy (mean value ± standard deviation, hereafter the same) and 9.7 ± 1.4 mGy, and the maximum spinal cord, esophagus and ribs doses were 17.6 ± 4.2 mGy, 17.1 ± 3.3 mGy and 37.8 ± 5.3 mGy, respectively. For an in-treatment 4D-CBCT scan, the dose parameters were 5.6 ± 1.8 mGy, 5.6 ± 1.5 mGy, 10.3 ± 3.5 mGy, 10.3 ± 4.3 mGy and 22.4 ± 5.3 mGy, respectively. The OAR dose parameters for in-treatment 4D-CBCT scans showed a relatively strong linear relationship with beam MU and number of frames. For patients receiving 5-fraction treatments and one pre-treatment and one in-treatment 4D-CBCT for each fraction, the increases in lung V20, lung V5, mean lung and heart doses, maximum spinal cord, esophagus and ribs doses were 0.04 ± 0.03%, 0.29 ± 0.23%, 76.3 ± 14.4 mGy, 76.5 ± 12.0 mGy, 76.3 ± 20.2 mGy, 70.5 ± 13.1 mGy and 101.2 ± 20.5 mGy, respectively. The increase in OAR dose parameters was proportional to the number of fractions.ConclusionThe imaging dose of pre-treatment and in-treatment 4D-CBCT was calculated using a validated MC model. For 5-fraction treatments, the imaging dose will likely have minimal clinical impact on OAR dose parameters.
More Related Videos
11:38Voluntary Breath-hold Technique for Reducing Heart Dose in Left Breast Radiotherapy
Published on: July 3, 2014
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography