1-4-2: Evaluation of applied mechanical power to individual lungs in a simulator-based setting of one ventilator for
Lars-Olav Harnisch1, Christian Czock1, Matthew A Levin2
1Department of Anaesthesiology, University Medical Center, Georg-August-University, Göttingen, Germany.
Background:
The concept of ventilating multiple patients concurrently using a single ventilator has been proposed as a solution when the demand for ventilators surpasses the available supply. While the practicality of this approach has been established, a thorough evaluation of the risks involved has yet to be comprehensively addressed.
Methods:
Two circuits, a simple one (circuit-1) and another with an adjustable resistance valve (circuit-2), were evaluated within an experimental framework utilizing two computer-controlled lung simulators (TestChest and ASL 5000). These simulators were ventilated by an ICU (intensive care unit) ventilator (Servo-u) employing various ventilation modes (volume- and pressure-controlled ventilation). The study was conducted under differing respiratory conditions, characterized by low compliance (20 ml/cmH2O) as well as normal-high compliance (100 ml/cmH2O), in order to ascertain the applied tidal volume (VT), pressures, and the resultant mechanical power (MP).
Results:
Circuit-1: The applied VT, pressures, and MP differed significantly between the two simulators, as well as in relation to ventilation mode, compliance, and respiratory rate (RR) (p < 0.001); the differences were most pronounced in settings with differing compliance levels. Circuit-2: Differences in VT, pressures, and MP were observed between simulators concerning valve settings (p < 0.001). The VT demonstrated a negative correlation, with volumes derived from valve closure spanning from 50 to 100 ml across all settings. In the design of circuit-2, MP exceeded the 12 J/min threshold in both lung simulators at elevated RR and could only be decreased through valve closure followed by a consequential hypoventilation in one simulator.
Conclusion:
The simultaneous ventilation of two patients using a single ventilator is technically viable, yet it presents considerable risks. Even with the integration of an adjustable resistance valve to accommodate varying lung complexities, the likelihood of unilateral hypoventilation and elevated mechanical stress remains high.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Factors Affecting Pulmonary Ventilation
Alveolar Surface Tension
The alveolar fluid lines the luminal surface of the alveoli and exerts a force called surface tension. This force is caused by the polar water molecules in the liquid being more strongly attracted to each...
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation: