Biomechanical evaluation of bispherical augments in acetabular defect reconstruction and comparative study of
Xiaoqi Zhang1,2, Guoyuan Li2, Min Chen2
1Cheeloo College of Medicine, Shandong University, Jinan, 250012, Shandong, China.
Background:
Novel bispherical augments developed for acetabular defect reconstruction, which combined with acetabular components, have shown excellent results in clinical follow-up. This study aims to evaluate the primary stability of them in severe acetabular defect reconstruction and biomechanical effect using different combine methods in the component-augment interface, and to further explain their value in clinical applications.
Methods:
Nine composite hemipelvis specimens were prepared with a Paprosky II defect. A porous titanium acetabular shell and bispherical augment were combined and implanted for acetabular defect reconstruction. The augment was fixed to the hemipelvis using screws, while three specimens per groups of three groups were varied by different fixation techniques between the augments and acetabular components: (1) screw fixation only, (2) cement fixation only, and (3) screw and cement fixation. The combined components and hemi pelvic specimens were cyclically loaded (500 cycles) with three different levels of loads (maximum 0.5 kN, 0.9 kN, and 1.8 kN).
Results:
All constructs had micromotions of less than 99 μm, and bony ingrowth could be achieved between the prostheses and the bone in 3-30% and 5-50% loads. The cement groups showed the maximum relative movement in the interface between the bone and combined prosthesis, especially when the load was increased to 1.8 kN (100%), while in which the hybrid groups had the minimum value.
Conclusion:
The study indicates that acetabular defect reconstruction utilizing the bispherical augments can achieve excellent primary stability, which provides the mechanical condition of bone ingrowth for prostheses and augments. The hybrid fixation is the most reliable of them, even under full loads.
More Related Videos
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
08:40Author Spotlight: A Novel 3D-Printed Titanium Implant for Minimally Invasive Treatment of Hip Dysplasia in Young Dogs
Published on: April 19, 2024
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the procedure...
Bone Remodeling and Repair