Related Experiment Video
Updated: Jul 1, 2026

Dependence of Laser-induced Breakdown Spectroscopy Results on Pulse Energies and Timing Parameters Using Soil Simulants
Published on: September 23, 2013
Multiple laser parametric study of fiber-laser-induced breakdown spectroscopy
Abstract:
An integrated investigation of pulse energy (0-0.8 mJ), pulse width (2-500 ns), and repetition rate (1 Hz-4 MHz) was carried out on laser-induced breakdown spectrometry (LIBS). An increase in pulse energy led to a significant signal elevation, yet its speed slowed down at a longer pulse width, which helps to lessen the electron density and ablation threshold. The repetition rate showed greater impact at higher pulse energy, and a threshold of about 200 Hz was observed for signal enhancement, exceeding which the ablation threshold of aluminum decreases logarithmically in the case of a 250 ns pulse width. Up to 29.0% of the sensitivity enhancement and 43.9% of the limit of detection (LOD) improvement were obtained under a high laser repetition rate of 20 kHz, resulting in an agreeable LOD of down to several tens of ppm considering the low laser irradiance applied here. In summary, to achieve better quantification results, higher pulse energy, moderate pulse width, and higher repetition rate should be applied.
More Related Videos
09:40Measurement and Analysis of Atomic Hydrogen and Diatomic Molecular AlO, C2, CN, and TiO Spectra Following Laser-induced Optical Breakdown
Published on: February 14, 2014
10:17Laser-induced Breakdown Spectroscopy: A New Approach for Nanoparticle's Mapping and Quantification in Organ Tissue
Published on: June 18, 2014
Related Concept Videos
MALDI-TOF Mass Spectrometry
Mass Spectrum: Interpretation