A Method for Quantifying Upper Limb Performance in Daily Life Using Accelerometers
Published on: April 21, 2017
Comparison of ActiGraph CentrePoint Insight Watch Placement on Dominant and Nondominant Wrists in Young Adults in
Daehyoung Lee1, Haley Voermans-Dean2,3, Jung Eun Lee2
1Department of Health Behavior and Nutrition Sciences, University of Delaware, 26 North College Avenue (013 Carpenter Sports Building), Newark, Delaware, 19716, United States, 1 302 831 0762.
Background:
With the continuous evolution of technology, wearable accelerometers have become one of the most popular means of measuring daily physical activity (PA) levels. Despite the conventional use of the nondominant wrist as a device placement in numerous PA studies, the impact of wrist-worn accelerometer placement on PA data outcomes remains uncertain.
Objective:
This study aimed to examine the degree of agreement between accelerometry data collected from CentrePoint Insight Watches (CPIWs; ActiGraph) worn on the dominant and nondominant wrists of young adults in free-living conditions.
Methods:
Twenty-nine participants (mean age 20.2, SD 1.6 years; 23 females) simultaneously wore an ActiGraph CPIW on both dominant and nondominant wrists for 7 consecutive days during waking hours. A sampling frequency of 32 Hz and Montoye 2020 cut-points were used to categorize activity intensity based on counts per minute. Data validity criteria included (1) ≥600 minutes per day of monitor wear time for both wrists, (2) a daily wear time difference of <1% of the average wear time between the dominant and nondominant wrists, and (3) a minimum of 3 valid days of monitor wear for both wrists. Bland-Altman plots and intraclass correlation coefficient (ICC) analyses were performed to compare the accelerometry data between the two device placements.
Results:
Average daily monitor wear time was 789.6 (SD 86.1) minutes per day for the dominant wrist and 793.0 (SD 91.8) minutes per day for the nondominant wrist. All accelerometer variables, including sedentary time (ST), light PA, moderate-to-vigorous PA (MVPA), steps, triaxial counts, and vector magnitude (VM), showed good-to-excellent levels of reliability between the two measurements (ICC >0.88 for all; P<.001). Bland-Altman analysis calculated mean bias and SD between the two device placements as follows: ST (-18.8, SD 27.6 min/d), light PA (2.7, SD 15.9 min/d), MVPA (12.7, SD 26.7 min/d), steps (218.1, SD 476.6 counts/d), x-axis (99.4, SD 188.8 counts/min), y-axis (73.9, SD 147.0 counts/min), z-axis (107.6, SD 183.5 counts/min), and VM (161.2, SD 273.4 counts/min). Bland-Altman plots revealed that the upper and lower limits of agreement across most variables were considerably wide.
Conclusions:
Our findings partially align with previous research, demonstrating higher MVPA and step counts on the dominant wrist, while the nondominant wrist produced a higher level of ST. Despite the acceptable level of reliability between the two placements based on ICC analyses, the dominant wrist tended to produce greater outcomes as the intensity of PA increased, highlighting the need for careful consideration when determining the wear location of CPIWs and interpreting data outcomes.
More Related Videos
05:51Assessing the Accuracy of Fitness Smartwatch Data for Cardiovascular and Physical Activity Monitoring: A Validation Study in Digital Health
Published on: February 21, 2025
05:03Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Assessment of radial pulse
The radial pulse, located at the wrist, is often the preferred site for assessing peripheral pulse because of its accessibility and dependability. The process of determining the radial pulse involves several steps:
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement