Disruption of Frontal Lobe Neural Synchrony During Cognitive Control by Alcohol Intoxication
Published on: February 6, 2019
From analytical interference to the detection of intoxication
Sonali Suresh Karnani-Khemlani1, María José de la Vega-Prieto1, Juana Reyes Padrón-Morales1
1Department of Laboratory Medicine, Complejo Hospitalario Universitario de Canarias, San Cristóbal de La Laguna, Spain.
Abstract:
This article describes the case of a 66-year-old man who attended the emergency department due to a decrease in the level of consciousness. When determining creatinine in blood plasma using the enzymatic method on the Roche cobas c702® (Roche Diagnostics, Manheim, Germany), an undetectable creatinine value was obtained. Recently, both the literature and the manufacturer of the assay have described that the toxic metabolite of paracetamol, N-acetyl-p-benzoquinone imine, interferes with the enzymatic measurement of creatinine. Therefore, the concentration of paracetamol was determined in the same sample confirming toxic levels. On the other hand, the determination of creatinine by the Jaffe method did not present such interference. This case underlines the importance of determining creatinine by non-enzymatic methods in cases of paracetamol overdose in order to correctly assess renal function.
More Related Videos
10:11Microdialysis of Ethanol During Operant Ethanol Self-administration and Ethanol Determination by Gas Chromatography
Published on: September 5, 2012
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Atomic Emission Spectroscopy: Interference
High-Performance Liquid Chromatography: Types of Detectors
Difference from Background: Limit of Detection
The LOD indicates the presence or absence...
Prevention of Further Absorption of Poison