G B Silva1, V R Botelho1, M S Alva-Sanchez1
1Graduate Program in Information Technology and Healthcare Management, Federal University of Health Sciences of Porto Alegre (UFCSPA), Sarmento Leite, 245, Porto Alegre, 90050-170, Brazil.
Background:
Reconstructing the X-ray spectrum is crucial for radiodiagnostic applications. The indirect method, based on the Archer & Wagner model, provides an approach to estimating X-ray spectra, but its accuracy and limitations require further investigation.
Objective:
This study aims to evaluate the effectiveness of the Archer & Wagner model in reconstructing Xray spectra by evaluating its performance with experimental data and simulations. Furthermore, the study examines the impact of a filtering technique introduced in a subsequent work by Archer & Wagner.
Methods:
The method was analyzed using experimental data and SpekPy simulations, alongside the generation of over 20,000 transmission curves and corresponding spectra from the Archer & Wagner model. Additionally, we assessed how Archer and Wagner's (1989) 0-3 g cm⁻² truncation and renormalization filter influences the accuracy of the indirectly estimated spectral parameters.
Results:
Findings indicate that, despite the model's strong ability to represent X-ray spectra, minimizing the discrepancy between transmission curves is not the most effective approach for parameter estimation. Additionally, the accuracy of X-ray parameters, such as the half-value layer and homogeneity coefficient, is closely linked to the quality of the reconstructed spectra. While the filtering technique enhances estimation, defining a standard cutoff remains a challenge.
Conclusion:
The Archer & Wagner model effectively represents X-ray spectra, but the optimization approach based on transmission curve minimization may not be the best for parameter estimation. The quality of reconstructed spectra directly influences key X-ray parameters, and further refinement of filtering techniques is needed to establish a reliable standard cut-off.
More Related Videos
06:49In situ Grazing Incidence Small Angle X-ray Scattering on Roll-To-Roll Coating of Organic Solar Cells with Laboratory X-ray Instrumentation
Published on: March 2, 2021
07:55Elemental-sensitive Detection of the Chemistry in Batteries through Soft X-ray Absorption Spectroscopy and Resonant Inelastic X-ray Scattering
Published on: April 17, 2018
Related Concept Videos
Spectrophotometry: Introduction
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
Atomic Emission Spectroscopy: Lab
UV–Vis Spectrometers
Atomic Emission Spectroscopy: Overview
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
Atomic Emission Spectroscopy: Instrumentation