Immersion factor determination for TriOS RAMSES radiance sensors
Optics Express
|August 13, 2025
Summary
Immersion factors for TriOS RAMSES radiance sensors were determined, showing minimal variation between sensors and previous studies. A consistent average immersion factor is proposed for the RAMSES sensor class.
Area of Science:
- Ocean optics
- Radiometry
- Sensor calibration
Background:
- Accurate in-water radiance measurements are crucial for oceanographic studies.
- Immersion factors are essential for correcting sensor readings in aquatic environments.
- Previous studies established immersion factors for RAMSES hyper-spectral radiometers.
Purpose of the Study:
- Determine immersion factors for TriOS RAMSES ARC series radiance sensors.
- Compare these factors with existing data for the same sensor class.
- Propose a representative immersion factor for the RAMSES radiance sensor class.
Main Methods:
- Adapted a method from Zibordi (2006) to determine immersion factors.
- Utilized two distinct light sources: LED and tungsten-halogen lamps.
- Compared results with Zibordi and Darecki (2006) for RAMSES hyper-spectral radiometers.
Main Results:
- Immersion factors for the two tested sensors differed by 0.2%.
- The determined factors differed by an average of 0.3% from Zibordi and Darecki (2006).
- Results indicate consistency in immersion factors across the RAMSES radiance sensor class.
Conclusions:
- Immersion factors for RAMSES radiance sensors are sufficiently constant for class representation.
- An average immersion factor, derived from this study and prior work, is proposed.
- The proposed average immersion factor shows a root-mean-square-difference <0.2% from 400 nm to 900 nm.
Related Concept Videos
Impedance Combination
504
Consider a string of christmas lights, each bulb symbolizing an impedance element. In this series configuration, the flow of electric current remains uniform across every component. This behavior aligns with Kirchhoff's Voltage Law (KVL), which asserts that the total impedance in such a setup equals the sum of individual impedances—akin to resistors in series. It follows that the voltage from the power source is distributed proportionally among these components, adhering to the...
504
Raman Spectroscopy Instrumentation: Overview
532
A conventional Raman spectrophotometer includes a laser source, a sample holding system, a wavelength selector, and a detector.
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
532
Mass Analyzers: Common Types
705
The quadrupole mass analyzer consists of four cylindrical metal rods arranged in a diamond carrying a DC voltage and a radio-frequency AC voltage. The motion of ions through the quadrupole depends on the field strength, causing only ions of a certain m/z to resonate successfully and strike the detector at a given field strength. Though the transmission rate for these analyzers is high, the exact elemental composition of the sample is not determined because of low resolution; however, they are...
705