Analytical Performance of the New Sysmex High-Sensitivity Troponin T Assay
Chin Shern Lau1, Napthathorn Asavapuriyothin2, Chee Hoe Low2
1Department of Laboratory Medicine, Changi General Hospital, Singapore 529889, Singapore.
The Sysmex HISCL hs-cTnT assay is a high-sensitivity troponin assay that performs comparably to the Roche assay. Baseline troponin levels increase with age and decreased kidney function.
Area of Science:
- Clinical Chemistry
- Cardiovascular Diagnostics
- Immunoassay Technology
Background:
- High-sensitivity cardiac troponin T (hs-cTnT) assays are crucial for diagnosing acute myocardial infarction.
- Evaluating new diagnostic platforms like the Sysmex Highly Integrated Single-Cartridge Luminescence Immunoassay System (HISCL) is essential for clinical practice.
- Comparison with established assays and determination of reference limits are key performance indicators.
Purpose of the Study:
- To evaluate the Sysmex HISCL hs-cTnT assay's analytical performance.
- To compare its performance against the Roche hs-cTnT assay.
- To establish 99th-percentile upper reference limits (99% URLs) in a healthy population and assess the impact of age and estimated glomerular filtration rate (eGFR).
Main Methods:
- Assay verification included limits of blank/detection, precision, and functional sensitivity.
- Method comparison was performed using linear regression and Bland-Altman analysis between Sysmex HISCL and Roche Elecsys analyzers.
- 99% URLs were derived from a healthy cohort (n=1004), and the influence of age and eGFR on hs-cTnT levels was analyzed.
Main Results:
- The HISCL assay demonstrated low limits of detection (1.3 ng/L) and blank (1.9 ng/L), with good precision (CVs < 10% at 99% URLs).
- Excellent agreement was observed with the Roche assay (r=0.95), with minimal mean differences.
- The 99% URLs were established as 14.4 ng/L (overall), 17.0 ng/L (males), and 13.9 ng/L (females), with hs-cTnT levels increasing with age and decreasing eGFR.
Conclusions:
- The Sysmex HISCL hs-cTnT assay meets the criteria for a high-sensitivity assay.
- It demonstrates reliable performance and good agreement with a comparator assay.
- Established 99% URLs and observed age/eGFR effects are important considerations for clinical interpretation.
More Related Videos
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
09:23Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Related Concept Videos
Acute Coronary Syndrome III: Diagnostic Studies
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome I: Introduction