Related Experiment Video
Updated: Sep 11, 2025

The Cutting and Floating Method for Paraffin-embedded Tissue for Sectioning
Published on: September 5, 2018
Paraffin sectioning in the modern laboratory and the problem of static electricity: a review
Ronald Sluys1, Sytske de Waart1
1Naturalis Biodiversity Center, Leiden, The Netherlands.
Abstract:
Build-up of static electricity frequently hampers proper production and handling of serial paraffin sections produced on a microtome, a problem that is exacerbated in modern, fully climatized laboratories. In this context, we review a number of suggested remedies for static electricity, concerning increase of humidity, in one way or the other, and various kinds of anti-static devices. An excellent solution, the shockless one-point ionizing bar that we have implemented in our laboratory, albeit not new, is presented here in detail.
Related Concept Videos
Preparation of Samples for Electron Microscopy
Fixation and Sectioning
The simplest type of preparation is the wet mount, in which the specimen is placed in a drop of liquid on the slide. A liquid specimen can be directly deposited on the slide using a dropper. Solid specimens, such as skin scraping, can be placed on the slide before adding a drop of liquid to prepare the wet mount. Sometimes the liquid is simply water, but stains are often added...